FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® GLASS PLASMA TUBE LT. BLUE BD HEMOGARD¿

MDR report key: 7115162 · Received December 13, 2017

Report

Report Number
1917413-2017-00606
Event Type
Malfunction
Date Received
December 13, 2017
Date of Event
December 28, 2016
Report Date
December 12, 2017
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903697149
PMA / PMN Number
K945952
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: BD PAS RECEIVED(B)(4) SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE CUSTOMER SAMPLES WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE OF ¿DRAW VOLUME - UNDERFILL¿ AND "SPLIT TUBES" WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. WATER FILL TESTING WAS CONDUCTED ON THE CUSTOMER SAMPLES AND NO DEFECTS RELATED TO DRAW VOLUME WERE OBSERVED. THE MANUFACTURING RECORDS WERE REVIEWED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED. CONCLUSION: BD WAS UNABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE CUSTOMER SAMPLES. ONE ROOT CAUSE OF THE DAMAGE THAT IS BETWEEN STOPPER AND THE LABEL IS A J CRACK. IT CAN HAPPEN DURING THE FORMING PROCESS. THIS OCCURS AT THE POINT WHERE THE GLASS IS BEING HEATED TO BE CUT. THE REASON THAT J CRACKS OCCUR IS DUE TO THE QUALITY OF THE CANE GLASS OR IF THE HEAT IS TOO HIGH. IF THE TUBE GETS A J CRACK WHEN THE STOPPER/SHIELD IS ASSEMBLED ONTO THE TUBE IT CAN CAUSE THE TUBE TO BREAK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD VACUTAINER® GLASS PLASMA TUBE LT. BLUE BD HEMOGARD¿ TUBES HAVE STOPPED FILLING AND TWO HAVE SPLIT IN THE AREA UNDER THE CAP AND ABOVE THE TOP OF THE LABEL. THE TUBES ARE VERY COLD TO TOUCH AND HAVE A LARGER AMOUNT OF "FOGGY AREA" INSIDE THEM. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
891049 BD VACUTAINER® GLASS PLASMA TUBE LT. BLUE BD HEMOGARD¿ BLOOD COLLECTION TUBE JKA BECTON, DICKINSON & CO. 5272777 50382903697149

Patients

Seq Age Sex Outcome Treatment
1 Other