FDA Adverse Event Malfunction Summary report: N

VISUAL-ICE® CRYOABLATION SYSTEM

MDR report key: 7111217 · Received December 12, 2017

Report

Report Number
3004462490-2017-00010
Event Type
Malfunction
Date Received
December 12, 2017
Date of Event
November 8, 2017
Report Date
December 12, 2017
Manufacturer
GALIL MEDICAL INC.
Product Code
GEH
UDI-DI
00867379000006
PMA / PMN Number
K113860
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

GALIL MEDICAL TECH SUPPORT IN ISRAEL DISCUSSED THE ISSUE WITH THE CUSTOMER AND DETERMINED THE ISSUE WAS RELATED TO THE PM PARTS. GALIL MEDICAL SENT A NEW PM KIT TO THE CUSTOMER. THE CUSTOMER INSTALLED THE PM KIT. FOLLOW-UP TESTING SHOWED THE SYSTEM PERFORMED AS EXPECTED. THE FAILED PM KIT WAS RETURNED TO GALIL FOR FURTHER INVESTIGATION. INVESTIGATION SHOWED DESICCANT DUST CLOGGING THE TUBE OUTLET FILTER CAUSING LOW GAS FLOW THROUGH MOST CHANNELS RESULTING IN SMALL ICE FORMATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT WHEN THE PHYSICIAN CHECKED THE ICE-BALL SIZE DURING PROCEDURE THROUGH ULTRASOUND, THE ICE-BALL SIZE SEEMED A LOT SMALLER THAN USUAL. AFTER FINISHING THE PROCEDURE, THE PHYSICIAN AGAIN CHECKED THE ICE-BALL SIZE ON THE CT SCAN AND CONFIRMED THAT THE ICE-BALL SIZE WAS DISTINCTLY SMALL. THE PHYSICIAN REPORTED THAT THE PROCEDURE WAS NOT SUCCESSFUL, BECAUSE ICE-BALL WAS NOT MADE WELL. THE PATIENT WAS UNDER LOCAL ANESTHESIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
887573 VISUAL-ICE® CRYOABLATION SYSTEM CRYOSURGICAL DEVICES AND ACCESSORIES GEH GALIL MEDICAL INC. FPRCH6000 00867379000006

Patients

Seq Age Sex Outcome Treatment
1