FDA Adverse Event Malfunction Summary report: N

VITALITY AVT

MDR report key: 710795 · Received May 9, 2006

Report

Report Number
2124215-2006-13792
Event Type
Malfunction
Date Received
May 9, 2006
Date of Event
February 16, 2006
Report Date
February 16, 2006
Manufacturer
CARDIAC PACEMAKERS, INC
Product Code
LWS
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
LAY USER/PATIENT

Narratives

Description of Event or Problem · 1

EVENT DESCRIPTION GUIDANT RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD A CHARGE TIME OF 12 SECONDS 6 MONTHS AGO. IN MOL-1 WITH 2.64 V. CURRENTLY, THE CHARGE TIME IS 21.8 SEC. A FEW MANUAL CAPACITOR REFORMATIONS INCREASED THE CHARGE TIME TO 25 AND THE DEVICE DISPLAYED AN ELECTIVE REPLACEMENT INDICATOR (ERI).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY AVT IMPLANTABLE ATRIAL GENERATOR LWS CARDIAC PACEMAKERS, INC A135 NA

Patients

Seq Age Sex Outcome Treatment
1 87 YR Other