FDA Adverse Event
Malfunction
Summary report: N
VITALITY AVT
MDR report key: 710795
·
Received May 9, 2006
Report
- Report Number
- 2124215-2006-13792
- Event Type
- Malfunction
- Date Received
- May 9, 2006
- Date of Event
- February 16, 2006
- Report Date
- February 16, 2006
- Manufacturer
- CARDIAC PACEMAKERS, INC
- Product Code
- LWS
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- LAY USER/PATIENT
Narratives
Description of Event or Problem · 1
EVENT DESCRIPTION GUIDANT RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD A CHARGE TIME OF 12 SECONDS 6 MONTHS AGO. IN MOL-1 WITH 2.64 V. CURRENTLY, THE CHARGE TIME IS 21.8 SEC. A FEW MANUAL CAPACITOR REFORMATIONS INCREASED THE CHARGE TIME TO 25 AND THE DEVICE DISPLAYED AN ELECTIVE REPLACEMENT INDICATOR (ERI).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY AVT | IMPLANTABLE ATRIAL GENERATOR | LWS | CARDIAC PACEMAKERS, INC | A135 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Other |