FDA Adverse Event Malfunction Summary report: N

GOLDTRACE FETAL SCALP ELECTRODE

MDR report key: 7106746 · Received December 11, 2017

Report

Report Number
1722684-2017-00010
Event Type
Malfunction
Date Received
December 11, 2017
Date of Event
November 17, 2017
Report Date
November 12, 2018
Manufacturer
CLINICAL INNOVATIONS, LLC
Product Code
HGP
PMA / PMN Number
K030691
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL INNOVATIONS WAS UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS INCIDENT TO FURTHER EVALUATE THE CAUSE.

Description of Event or Problem · 0

HOPSITAL NARRATIVE TRANSLATED TO ENGLISH: "IN THE LAST TWO WEEKS WE HAD COMPLICATIONS WITH THE SCALP ELECTRODE : - 1X WINDINGS LEFT WITH INFECTION - 1X LACERATION OF THE SKIN AFTER REMOVING. IN GENERAL WE ALL HAVE THE IMPRESSION THAT THE SCALP ELECTRODES ARE VERY DIFFICULT TO DETACH. IT IS FOR EXAMPLE AFTER A CAESAREAN SECTION OFTEN NOT POSSIBLE TO REMOVE THE ELECTRODE BY PULLING APART THE THREADS. THE NURSES ALSO HAVE THE FEELING THAT THE SPIRAL IS LONGER THAN BEFORE. THE LOT NUMBER OF THESE ELECTRODES IS 170347. WE WOULD LIKE TO RECEIVE YOUR REACTION TO THIS. CAN YOU ALSO LET US KNOW IF IT IS CORRECT THAT THE SPIRAL IS NOW LONGER THAN IN THE PAST? DO YOU HAVE AN INSTRUCTION TO REMOVE CORRECTLY THE SCALP ELECTRODE? MAYBE THE CUSTOMER CAN LEARN FROM THAT FOR THE FUTURE."

Additional Manufacturer Narrative · 1

THIS HAS AN EXTREMELY RARE OCCURRENCE AND THIS TYPE OF FAILURE IS CAUSED BY USE ERROR OF THE PRODUCT EITHER WHEN PLACING THE DEVICE (OVER-ROTATION) AND/OR UPON REMOVING THE DEVICE (UNDER-ROTATION). CLINICAL INNOVATIONS IS TRYING TO GET MORE INFOMRATION REGARDING THE INCIDENT.

Description of Event or Problem · 1

HOSPITAL NARRATIVE TRANSLATED TO ENGLISH: "IN THE LAST TWO WEEKS WE HAD COMPLICATIONS WITH THE SCALP ELECTRODE: -1X WINDING LEFT WITH INFECTION - 1X LACERATION OF THE SKIN AFTER REMOVING. IN GENERAL WE ALL HAVE THE IMPRESSION THAT THE SCALP ELECTRODES ARE VERY DIFFICULT TO DETACH. IT IS FOR EXAMPLE AFTER A CAESAREAN SECTION OFTEN NOT POSSIBLE TO REMOVE THE ELECTRODE BY PULLING APART THE THREADS. THE NURSES ALSO HAVE THE FEELING THAT THE SPIRAL IS LONGER THAN BEFORE. THE LOT NUMBER OF THESE ELECTRODES IS 170347. WE WOULD LIKE TO RECEIVE YOUR REACTION TO THIS. CAN YOU ALSO LET US KNOW IF IT IS CORRECT THAT THE SPIRAL IS NOW LONGER THAN IN THE PAST? DO YOU HAVE AN INSTRUCTION TO REMOVE CORRECTLY THE SCALP ELECTRODE? MAYBE THE CUSTOMER CAN LEARN FROM THAT FOR THE FUTURE"

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
885519 GOLDTRACE FETAL SCALP ELECTRODE FETAL SPIRAL ELECTRODE HGP CLINICAL INNOVATIONS, LLC CNS00004 170347

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization