FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 1.0

MDR report key: 7106354 · Received December 11, 2017

Report

Report Number
3007042319-2017-04714
Event Type
Malfunction
Date Received
December 11, 2017
Date of Event
November 8, 2017
Report Date
August 20, 2019
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Removal / Correction Number
Z-1903-2018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY #2018-04-02, 11:31:26, HERRGC1: PRODUCT EVENT SUMMARY: THE CONTROLLER ((B)(6)) AND EIGHT BATTERIES ((B)(6)) WERE RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED BATTERIES REVEALED THAT THE DEVICES PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. THE CONTROLLER PASSED FUNCTIONAL TESTING; HOWEVER, VISUAL INSPECTION REVEALED LOOSE POWER PORT 1 AND 2 CONNECTORS. THIS IS AN ADDITIONAL OBSERVATION NOT RELATED TO THE REPORTED EVENT. BASED ON AN INTERNAL INVESTIGATION, THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO INADEQUATE THREAD LOCK, AN INCONSISTENT THREAD LOCK CURE TIME AND AN INADEQUATE TORQUE APPLICATION DURING THE ASSEMBLY PROCESS. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER CONTAINED THE SMR SOFTWARE. THE SOFTWARE UPGRADE CONTAINS A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(6). LOG FILES ANALYSIS ALSO REVEALED POWER SWITCHING EVENTS DUE TO COMMUNICATION ERROR S INVOLVING (B)(6). AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. MOMENTARY DISCONNECTIONS ARE BEING INTERNALLY INVESTIGATED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. ADDITIONAL PRODUCTS: BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 BATTERY (B)(6). H3: YES H6 FDA METHOD CODE(S): 10, 23, 3372, 38 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 25 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CORRECTIONS: B3, B5, H6, H7, H9, H10. PRODUCT EVENT SUMMARY: THE CONTROLLER (B)(6) AND EIGHT (8) BATTERIES (B)(6) WERE RETURNED FOR EVALUATION. A REVIEW OF THE (B)(6)' S MANUFACTURING DOCUMENTATION CONFIRMED THAT THE ASSOCIATED CONTROLLER MET ALL REQUIREMENTS FOR RELEASE. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED BATTERIES REVEALED THAT THE DEVICES PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. THE RETURNED CONTROLLER PASSED FUNCTIONAL TESTING. VISUAL INSPECTION REVEALED LOOSE POWER PORT ONE (1) AND TWO (2) CONNECTORS. THIS IS AN ADDITIONAL OBSERVATION NOT RELATED TO THE REPORTED EVENT. BASED ON AN INTERNAL INVESTIGATION CONDUCTED, THE MOST LIKELY ROOT CAUSE OF THE LOOSE CONNECTORS CAN BE ATTRIBUTED TO INADEQUATE THREAD LOCK, AN INCONSISTENT THREAD LOCK CURE TIME AND AN INADEQUATE TORQUE APPLICATION DURING THE ASSEMBLY PROCESS. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER IN USE DURING THE REPORTED EVENT, (B)(6), CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILE REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(6), AS WELL AS PREMATURE POWER SWITCHING EVENTS DUE TO COMMUNICATION ERRORS INVOLVING (B)(6). DATA LOG FILES ALSO REVEALED MULTIPLE INSTANCES INVOLVING (B)(6) WHERE THE BATTERY¿S RELATIVE STATE OF CHARGE (RSOC) WAS BETWEEN 101-20, WHICH ARE INDICATIVE OF COMMUNICATION ERRORS. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. POSSIBLE ROOT CAUSES OF THE COMMUNICATION ERRORS CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS ON THE COMMUNICATION PINS OF THE CONTROLLER, THE CONTROLLER NOT RECEIVING RESPONSES FROM THE BATTERY, AND/OR DUE TO THE PACKET ERROR CHECKING METHOD DETECTING BIT ERRORS. THE MOST LIKELY ROOT CAUSE OF THE REPORTED PREMATURE POWER SWITCHING CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. AN INTERNAL INVESTIGATION EXAMINED MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: D4: SERIAL#: (B)(6). H6: FDA METHOD CODE(S): 4112 THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

LOG FILE REVIEW INDICATED ONE BATTERY ALSO HAD A COMMUNICATION ERROR.

Additional Manufacturer Narrative · 1

OTHER DEVICES INVOLVED IN THIS EVENT: MEDICAL DEVICE : BATTERY / (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 12/31/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 12/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 11/30/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE:11/30/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 12/31/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 12/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 05/31/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 05/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 03/31/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 03/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-03-31 UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 03/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 03/31/2017, UDI #: (B)(4). DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 03/31/2016, LABELED FOR SINGLE USE: NO. (B)(4). BRAND NAME: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY: BATTERY /(B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 09/30/2017, UDI #: (B)(4) DEVICE AVAILABLE FOR EVALUATION: YES, RETURN DATE: 11/21/2017. DEVICE EVALUATED BY MANUFACTURER: NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. DEVICE MANUFACTURE DATE: MFG DATE: 09/30/2016, LABELED FOR SINGLE USE: NO. (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CONTROLLER AND BATTERIES EXHIBITED POWER SWITCHING, DESPITE BATTERY CAPACITIES THAT WERE GREATER THAN 25%. THE CONTROLLER AND BATTERIES WERE EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
884101 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 1.0 VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1407DE

Patients

Seq Age Sex Outcome Treatment
1 49 YR (B)(6)