FDA Adverse Event Injury Summary report: N

UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM

MDR report key: 7105395 · Received December 11, 2017

Report

Report Number
2523190-2017-00152
Event Type
Injury
Date Received
December 11, 2017
Report Date
December 6, 2017
Manufacturer
INTEGRA YORK, PA INC.
Product Code
FST
PMA / PMN Number
K864380
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR EVALUATION. THE ULTRALITE HEADLIGHT SYSTEM WAS RETURNED IN USED CONDITION WITH A REPLACEMENT CABLE ATTACHED. THE REPLACEMENT CABLE'S "Y" JUNCTION HAD BECOME SEPARATED ON ONE SIDE. THE HEADLIGHT WAS FOUND TO SHOW CUSTOMER LABELING AND WEAR. THE FIBER OPTIC CABLE WAS DISCOLORED, WORN, AND DAMAGED. EVALUATION OF THE DEVICE SHOWED NO MANUFACTURING, WORKMANSHIP, OR MATERIAL DEFICIENCY. A REVIEW OF MANUFACTURING RECORDS FOUND NO NONCONFORMANCES. CORRECTED DATA: THE INITIAL REPORTED INDICATED THE HEADLIGHT CABLE WAS LOOSE AND NOT SAFE FOR USE. THE INABILITY TO USE THE DEVICE RESULTED IN A HOUR DELAY IN A PATIENT PROCEDURE. (B)(4).

Additional Manufacturer Narrative · 1

TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.

Description of Event or Problem · 1

THE HEADLIGHT CABLE IS LOOSE AND IT IS NOT SAFE FOR USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
884057 UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM CABLES FST INTEGRA YORK, PA INC.

Patients

Seq Age Sex Outcome Treatment
1 Other