UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM
Report
- Report Number
- 2523190-2017-00152
- Event Type
- Injury
- Date Received
- December 11, 2017
- Report Date
- December 6, 2017
- Manufacturer
- INTEGRA YORK, PA INC.
- Product Code
- FST
- PMA / PMN Number
- K864380
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED FOR EVALUATION. THE ULTRALITE HEADLIGHT SYSTEM WAS RETURNED IN USED CONDITION WITH A REPLACEMENT CABLE ATTACHED. THE REPLACEMENT CABLE'S "Y" JUNCTION HAD BECOME SEPARATED ON ONE SIDE. THE HEADLIGHT WAS FOUND TO SHOW CUSTOMER LABELING AND WEAR. THE FIBER OPTIC CABLE WAS DISCOLORED, WORN, AND DAMAGED. EVALUATION OF THE DEVICE SHOWED NO MANUFACTURING, WORKMANSHIP, OR MATERIAL DEFICIENCY. A REVIEW OF MANUFACTURING RECORDS FOUND NO NONCONFORMANCES. CORRECTED DATA: THE INITIAL REPORTED INDICATED THE HEADLIGHT CABLE WAS LOOSE AND NOT SAFE FOR USE. THE INABILITY TO USE THE DEVICE RESULTED IN A HOUR DELAY IN A PATIENT PROCEDURE. (B)(4).
TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
THE HEADLIGHT CABLE IS LOOSE AND IT IS NOT SAFE FOR USE.