ABUT HEX-LOCK 3.5MM IMP 5 .5 FLARE
Report
- Report Number
- 0001038806-2017-00890
- Event Type
- Malfunction
- Date Received
- December 8, 2017
- Date of Event
- October 24, 2017
- Report Date
- January 18, 2018
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- PK011028
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONE HEX-LOCK ABUTMENT (HLA3/5) WAS RETURNED FOR INSPECTION. A VISUAL INSPECTION REVEALED SIGNS OF WEAR DUE TO USAGE AROUND THE HEX AND THE COLLAR. A DEVICE HISTORY REVIEW WAS PERFORMED AND NO RELATED NONCONFORMANCE¿S WERE NOTED. A COMPLAINT HISTORY SEARCH WAS PERFORMED USING OUR COMPLAINT HANDLING SYSTEM AND THERE WERE NO ADDITIONAL RELATED COMPLAINTS AGAINST THIS LOT NUMBER. APPROPRIATE DOCUMENTATION WAS REVIEWED AND THE FOLLOWING INFORMATION WAS IDENTIFIED: INSTRUCTIONS FOR USE FOR ZIMMER DENTAL IMPLANT SYSTEMS (4894I REV. 3 - 07/09). SEATING OF PROSTHETIC COMPONENTS INTERNAL HEX OR OCTAGON COMPONENTS - TO PROPERLY SEAT THESE PROSTHETIC COMPONENTS, PLACE THE ABUTMENT RETAINING SCREW THROUGH THE ABUTMENT AND PLACE THE ASSEMBLY INTO THE INTERNAL BEVEL AT THE CORONAL PORTION OF THE IMPLANT. ROTATE IT UNTIL THE ABUTMENT DROPS INTO PLACE. THE MALE HEX OR OCTAGON SHOULD SEAT FULLY IN THE FEMALE HEX OR OCTAGON OF THE IMPLANT ONCE THE TORQUE WRENCH HAS BEEN USED TO TIGHTEN THE ABUTMENT TO 30NCM. A SINGULAR CAUSE FOR THE COMPLAINT COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED AS THE EVENT CANNOT BE RECREATED. COMPLAINT IS THEREFORE NON-VERIFIABLE. THE FOLLOWING SECTIONS HAVE BEEN UPDATED. (B)(4).
ADDITIONAL 510K NUMBERS: K013227 AND K953101
IT WAS REPORTED THAT THE ABUTMENT LOOSENED. TOOTH LOCATION 19.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 881056 | ABUT HEX-LOCK 3.5MM IMP 5 .5 FLARE | ABUTMENT | DZE | ZIMMER DENTAL | 63260194 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |