FDA Adverse Event Injury Summary report: N

POWERED WHEELCHAIR SYSTEM

MDR report key: 7095552 · Received December 7, 2017

Report

Report Number
9615350-2017-00007
Event Type
Injury
Date Received
December 7, 2017
Report Date
December 7, 2017
Manufacturer
MOTION CONCEPTS LP
Product Code
ITI
UDI-DI
00754014710065
PMA / PMN Number
K150574
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2017, "BOLT THAT HOLDS ARM BROKE" WAS REPORTED TO MOTION CONCEPTS BY THE DEALER ((B)(4)). ON (B)(4) 2017, UPON REQUEST BY THE DEALER, MOTION CONCEPTS SENT OUT A REPLACEMENT PARTS ORDER CONSISTING OF A TRC2536 CANTILEVER ARMREST. ASSEMBLY INSTRUCTIONS WERE PROVIDED BY MOTION CONCEPTS WITH THE PARTS ORDERS TO COMPLETE THE INSTALLATION OF THE REPLACEMENT PARTS. FROM THE END USER REPORT (EMAIL ON (B)(4) 2017), IT APPEARS THAT ONLY A PORTION OF THE CANTILEVER ARMREST ASSEMBLY WAS REPLACED BY THE DEALER, NOT THE COMPLETE REPLACEMENT ASSEMBLY THAT WAS SHIPPED BY MOTION CONCEPTS TO THE DEALER ON (B)(4) 2017. ON (B)(4) 2017, THE DEALER WAS ADVISED THAT MOTION CONCEPTS WILL BE SENDING ANOTHER REPLACEMENT PARTS ORDER AND THE COMPLETE ARMREST ASSEMBLY MUST BE INSTALLED AS PER THE INSTRUCTIONS PROVIDED WITH THE ORDER.

Description of Event or Problem · 1

ON 09-NOV-2017, MOTION CONCEPTS WAS NOTIFIED BY (B)(4) (MOTION CONCEPTS IMPORTER) THAT ONE OF THEIR DEALER (B)(4) REPORTED THE FOLLOWING INCIDENT: "THE RIGHT CHAIR ARM FELL AGAIN ON CONSTANCE. THIS TIME IT LEFT A 2.5X.5 ABRASION AND BRUISE. THE ARM WAS REPLACED AS WELL AS THE M8 HEX HEAD BOLT. HOWEVER, THE LEAF SPRING WHICH HOLDS THE ARM IN UPWARD POSITION WAS NOT. WHEN THE BOLT SHEARED THE FIRST TIME, IT BENT AND FATIGUED THE SPRING, "SO THAT NO IT MAKES NO CONTACT WITH THE HEX HEAD BOLT." IT IS BENT AWAY FROM THE ARM. THE LEFT ARM IS INTACT AND WORKS. IN THAT CASE, THE BOLT HEAD LOCKS UNDER THE SPRING IN THE UPRIGHT POSITION PREVENTING THIS TYPE OF INJURY. THIS NEEDS IMMEDIATE REMEDIATION. ALSO, THERE REMAINS A CLIP SPRING MISSING ON THE BACK-CONTROL WIRE COVER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
877539 POWERED WHEELCHAIR SYSTEM WHEELCHAIR ITI MOTION CONCEPTS LP 45T-12L-RESR-MAXX 00754014710065

Patients

Seq Age Sex Outcome Treatment
1 Other