FDA Adverse Event Injury Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 7092560 · Received December 7, 2017

Report

Report Number
2182208-2017-02024
Event Type
Injury
Date Received
December 7, 2017
Date of Event
November 27, 2016
Report Date
February 8, 2018
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. MULTIPLE PATIENTS WERE NOTED IN THE ARTICLE; HOWEVER, A ONE TO ONE CORRELATION COULD NOT BE MADE WITH UNIQUE DEVICE SERIAL NUMBERS. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. POSSIBLE MODELS INCLUDE THE MAXIMO ICD. REQUEST FOR ADDITIONAL INFORMATION WILL BE MADE AND UPON RECEIPT A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY. REFERENCED ARTICLE:A REVIEW OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR FAILURES DURING RADIATION THERAPY IN THREE (B)(6) HOSPITALS. JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY. 2017; 61(4):517-521.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

A JOURNAL ARTICLE WAS REVIEWED WHICH CONTAINED INFORMATION REGARDING IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FAILURES DURING RADIATION THERAPY. MULTIPLE PATIENTS WERE NOTED IN THE ARTICLE; HOWEVER, A ONE TO ONE CORRELATION COULD NOT BE MADE WITH UNIQUE DEVICE SERIAL NUMBERS. THE ARTICLE REPORTS ONE PATIENT EXPERIENCED ONE POWER ON RESET (POR), FOLLOWED BY REPROGRAMMING OF THE DEVICE. DURING ANOTHER ROUND OF TREATMENT, THE ICD EXPERIENCED A SECOND POR. THE ICD WAS REPROGRAMMED AND REMAINED IN USE. FURTHER FOLLOW UP DID NOT YET YIELD ANY ADDITIONAL INFORMATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

FOLLOW UP INFORMATION RECEIVED FROM THE AUTHOR INDICATED THERE ARE NO ADDITIONAL COMPLAINTS AND NO INFORMATION ON PRODUCT SERIAL NUMBERS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
873043 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MEDTRONIC, INC. MDT-ICD

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention