FDA Adverse Event Injury Summary report: N

CM DRIVE ACQUA IMPL TI 3.5X10

MDR report key: 7086611 · Received December 6, 2017

Report

Report Number
3008261720-2017-06670
Event Type
Injury
Date Received
December 6, 2017
Date of Event
October 26, 2017
Report Date
December 6, 2017
Manufacturer
NEODENT - JJGC S.A.
Product Code
DZE
UDI-DI
07899878016830
PMA / PMN Number
K133592
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2015015. INSTRADENT USA, INC IS SUBMITTING THE REPORT ON BEHALF OF NEODENT - JJGC

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2015015. INSTRADENT USA, INC IS SUBMITTING THE REPORT ON BEHALF OF NEODENT - JJGC. AFTER THE EVALUATION WAS NOTICED THAT THE DENTIST REPORTED THAT THE IMPLANT WAS IMMEDIATELY INSTALLED IN AN ALVEOLUS WITH SIGNS OF INJURY/ INFECTION. IN ADDITION, HE USED THE PROCEDURE OF IMMEDIATE LOAD. CONSIDERING THE SURGICAL METHODOLOGY, IT WAS SEEN THAT THE DENTIST HAS USED LESS DRILLS THAN RECOMMENDED TO PERFORM THE IMPLANT INSTALLATION.

Description of Event or Problem · 0

THE CLINICIAN REPORTED THAT EIGHT MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 4, OSSEOINTEGRATION WAS STILL NOT OBTAINED. CLINICIAN DID NOT PERFORM AN IMMEDIATE PLACEMENT NOR WAS THERE EVIDENCE OF FENESTRATION. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

Description of Event or Problem · 0

THE CLINICIAN REPORTED THAT EIGHT MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 4, OSSEOINTEGRATION WAS STILL NOT OBTAINED. CLINICIAN DID NOT PERFORM AN IMMEDIATE PLACEMENT NOR WAS THERE EVIDENCE OF FENESTRATION. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867791 CM DRIVE ACQUA IMPL TI 3.5X10 ENDOSSEOUS DENTAL IMPLANT DZE NEODENT - JJGC S.A. 48588I 07899878016830

Patients

Seq Age Sex Outcome Treatment
1 90 YR Required Intervention ABUTMENT| ABUTMENT| ABUTMENT