120 ML BD VACUTAINER® PLASTIC URINE COLLECTION CUP
Report
- Report Number
- 1917413-2017-00201
- Event Type
- Malfunction
- Date Received
- December 5, 2017
- Date of Event
- April 4, 2016
- Report Date
- January 24, 2018
- Manufacturer
- BECTON, DICKINSON & CO.
- Product Code
- KDT
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE INITIAL MDR WAS SUBMITTED WITH THE 510K NUMBER LISTED AS K790366. IT IS CORRECTED TO READ EXEMPT.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD 04/04/2016. RESULTS: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR FOREIGN MATTER OR MOLD ON URINE CUPS WITH THE INCIDENT LOT WAS NOT OBSERVED. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT# 5182571 AND NO ISSUES WERE IDENTIFIED. CONCLUSION: UNCONFIRMED COMPLAINT. AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMERS¿ INDICATED FAILURE MODE.
IT WAS REPORTED THAT THERE WAS A FOREIGN MATTER OR MOLD APPEARANCE ON 120 ML BD VACUTAINER® PLASTIC URINE COLLECTION CUPS. NO SERIOUS INJURY, BLOOD TO MUCOUS MEMBRANE EXPOSURE OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 864036 | 120 ML BD VACUTAINER® PLASTIC URINE COLLECTION CUP | URINE COLLECTION CUP | KDT | BECTON, DICKINSON & CO. | 5182571 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |