FDA Adverse Event Other Summary report: N

OLSEN MEDICAL

MDR report key: 708159 · Received April 18, 2006

Report

Report Number
MW1038729
Event Type
Other
Date Received
April 18, 2006
Date of Event
April 13, 2006
Report Date
April 18, 2006
Manufacturer
OLSEN MEDICAL
Product Code
GCJ
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

A BOVIE WAS USED OT MAKE AN INCISION ABOUT THE CIRCUMFERENCE OF THE TONSIL. DURING THE INCISION, A BURN WAS NOTED IN THE R LATERAL COMMISSURE. EXAMINE WITH THE BOVIE HANDPIECE REVEALED VIOLATION OF THE INSULATION OF THE BOVIE TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLSEN MEDICAL CAUTERY - BLUNT TIP SHORT GCJ OLSEN MEDICAL 66703 113741

Patients

Seq Age Sex Outcome Treatment
1 6 YR Other