FDA Adverse Event Malfunction Summary report: N

OAE PEDIATRIC FOAM EAR TIPS

MDR report key: 7081396 · Received December 4, 2017

Report

Report Number
3018859-2017-01003
Event Type
Malfunction
Date Received
December 4, 2017
Report Date
November 9, 2017
Manufacturer
NATUS MEDICAL INC.
Product Code
GWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS ISSUE WAS INVESTIGATED THROUGH CORRECTIVE ACTIONS. THE ROOT CAUSE OF THIS ISSUE WAS CONFIRMED TO BE INADEQUATE APPLICATION OF THE ADHESIVE WHICH SECURES THE FOAM TO THE LUMEN TUBE.

Additional Manufacturer Narrative · 1

THE CUSTOMER WAS PROVIDED REPLACEMENT FOAM TIPS. FOUR BAGS OF OAE PEDIATRIC FOAM TIPS WERE RETURNED TO NATUS. A TOTAL OF FOUR FOAM TIPS WERE TESTED. UPON VISUAL INSPECTION, ALL OF THE FOAM TIPS ARE SECURELY GLUED TO THE TUBE. SIX OUT OF THE TWELVE FOAM TIPS HAD SEPARATION FROM THE INNER TUBING WHEN PRESSURE WAS APPLIED. NATUS INSTRUCTIONS INCLUDE SELECTING THE APPROPRIATE SIZE EAR TIP FOR THE PATIENT, AS WELL AS INSTRUCTIONS REGARDING USING LARGER SIZE FOAM TIPS FOR INSERTION INTO THE EAR CANAL. NATUS IS INVESTIGATING THIS ISSUE THROUGH CORRECTIVE ACTIONS AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO NATUS MEDICAL THAT THE FOAM TIPS OF THE PEDIATRIC FOAM EAR TIPS ARE SEPARATING FROM THE INNER PLASTIC TUBING. THERE HAS BEEN NO REPORT OF PATIENT DEATH, INJURY OR DELAY IN PATIENT TREATMENT. NO ENVIRONMENTAL OR SAFETY CONCERNS HAVE ALSO BEEN REPORTED.