Report
- Report Number
- 1045254-2017-00430
- Event Type
- Malfunction
- Date Received
- November 29, 2017
- Date of Event
- November 14, 2017
- Report Date
- January 29, 2018
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- EQJ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT ANALYSIS INDICATES 1 OPENED AND 2 SEALED SAMPLES, PART NUMBER 1883673HS, FROM LOT NUMBER 0210640703 WERE RECEIVED. (ANALYSIS OF THE OPENED SAMPLE) THE INNER SHAFT BROKE 0.53¿ FROM THE DISTAL FACE OF THE INNER HUB WHICH WOULD HAVE RESULTED IN THE REPORTED MALFUNCTION. THE BREAK POINT CORRESPONDS TO THE PROXIMAL END OF THE OUTER TUBE IN THE FRONT HUB. THERE WERE STRIATIONS AROUND THE OUTSIDE DIAMETER OF THE BREAK POINT INDICATING METAL ON METAL CONTACT. WHEN VIEWED UNDER MAGNIFICATION, THERE WAS DAMAGE TO THE HUBS THAT IS CONSISTENT WITH IMPROPER LOADING: DIMPLES ON THE FRONT HUB PRIOR TO THE LOCKING AREA CAUSED BY THE HANDPIECE LOCKING MECHANISM; LOCKING AREA DEFORMATION CAUSED BY THE BACK SIDE OF THE FRONT COLLET OF THE HANDPIECE; AND DEFORMATION OF THE INNER HUB CHEVRONS CAUSED BY THE HANDPIECE DRIVE MECHANISM. THE INFORMATION INDICATES AN IMPROPERLY LOADED BUR CAUSED THE DEFORMATION OF THE HUBS; WHICH THEN CAUSED THE INNER SHAFT AND OUTER TUBE TO RUB TOGETHER UNTIL THE INNER SHAFT BROKE. THERE WAS NO INDICATION OF DEVICE FRAGMENTS AND THE BREAKAGE WOULD HAVE BEEN CONTAINED BY THE OUTER TUBE AND THE HANDPIECE. A BUR OR BLADE THAT IS NOT PROPERLY LOADED WOULD RESULT IN IMPROPER ENGAGEMENT WITH THE DRIVE MECHANISM OF THE HANDPIECE AND INSTABILITY OF THE BUR / BLADE. ONE OF THE SEALED SAMPLES WERE OPENED FOR THE ANALYSIS. THE BUR LOADED INTO A HANDPIECE, RAN AT 12,000 RPM IN FORWARD DIRECTION, AND CUT SAW BONE WITH NO ISSUES OR REPRODUCING ANY OF THE OBSERVE DAMAGE SEEN ON THE OPEN SAMPLE ABOVE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE CUSTOMER REPORTED THE BUR WAS FALLING APART AFTER A FEW MINUTES OF USE. IT WAS CONFIRMED THAT THE CASE WAS COMPLETED "SUCCESSFULLY BY USING A DIFFERENT BUR."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 847335 | XPS® BUR | BUR, EAR, NOSE AND THROAT | EQJ | MEDTRONIC XOMED INC. | 1883673HS | 0210640703 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |