FDA Adverse Event Malfunction Summary report: N

METRIX AIR

MDR report key: 7067196 · Received November 29, 2017

Report

Report Number
1000113657-2017-02000
Event Type
Malfunction
Date Received
November 29, 2017
Date of Event
November 6, 2017
Report Date
November 29, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00311917173894
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: (B)(4) -USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP UDI# (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE RESULTS. THE CUSTOMER IS CONCERNED WITH ALL OF THE RESULTS OBTAINED. THE EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE PRODUCT IS STORED BY CUSTOMER ACCORDING TO SPECIFICATION IN THE BEDROOM. DURING THE CALL ON DATE, A BACK TO BACK BLOOD TEST WAS PERFORMED NONFASTING BY THE CUSTOMER AND PRODUCED TEST RESULTS OF 280 AND 323MG/DL USING TRUE METRIX AIR METER. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/15/1208 AND OPEN VIAL DATE IS ONE MONTH AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (DATE/TIME NOT STORED CORRECTLY) RESULT 1 :202 MG/DL DATE: (B)(6) TIME:10:06 AM FASTING. RESULT 2 :266 MG/DL DATE: (B)(6) TIME:10:20 PM FASTING. RESULT 3 :219 MG/DL DATE: (B)(6) TIME:12:18 PM FASTING. RESULT 4 :200 MG/DL DATE: (B)(6) TIME:10:08 PM FASTING. RESULT 5 :241 MG/DL DATE: (B)(6) TIME:10:43 PM FASTING. CUSTOMER STATES THAT WHEN TESTING, HIS RESULTS ARE HIGHER THAN NORMAL FASTING. NO CHANGES TO EXERCISE AND/OR DIET ROUTINE. DATE AND TIME NOT CURRENTLY SET UP IN THE METER CORRECTLY. CUSTOMER IS CONCERNED WITH THE HIGH RESULTS FROM THE METER AND THE CUSTOMER IS ALSO CONCERNED WITH THE VARIANCE IN THE RESULTS FROM THE BACK TO BACK BLOOD TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
846383 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MU2515 00311917173894

Patients

Seq Age Sex Outcome Treatment
1 0 YR