FDA Adverse Event Malfunction Summary report: N

BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER.

MDR report key: 7066604 · Received November 29, 2017

Report

Report Number
1024879-2017-00656
Event Type
Malfunction
Date Received
November 29, 2017
Date of Event
February 4, 2016
Report Date
October 26, 2017
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686075
PMA / PMN Number
K982541
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. RESULTS: A SAMPLE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 5182956. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER SAFETY SHIELD CAME OFF WHEN TRYING TO CLOSE/ENGAGE. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
846023 BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER. SAFETY ENGINEERED SAMPLE NEEDLE FMI BECTON, DICKINSON & CO., (BD) 5182956 50382903686075

Patients

Seq Age Sex Outcome Treatment
1 Other