FDA Adverse Event Malfunction Summary report: N

7.5CM ANGLE ATTACHMENT

MDR report key: 7063585 · Received November 28, 2017

Report

Report Number
1045834-2017-50034
Event Type
Malfunction
Date Received
November 28, 2017
Date of Event
October 30, 2017
Report Date
October 30, 2017
Manufacturer
THE ANSPACH EFFORT LLC
Product Code
HBC
UDI-DI
00845384007159
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE OVERHEATING AT THE BEARINGS IN THE TIP 119 DEGREES. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. IT WAS ALSO NOTED THAT THE ROOT CAUSE OF THE HEAT WAS FROM WORN OUT BEARINGS AND THE COMPONENT DAMAGE WAS CAUSED FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS HEATING UP. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.