7.5CM ANGLE ATTACHMENT
Report
- Report Number
- 1045834-2017-50034
- Event Type
- Malfunction
- Date Received
- November 28, 2017
- Date of Event
- October 30, 2017
- Report Date
- October 30, 2017
- Manufacturer
- THE ANSPACH EFFORT LLC
- Product Code
- HBC
- UDI-DI
- 00845384007159
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE OVERHEATING AT THE BEARINGS IN THE TIP 119 DEGREES. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. IT WAS ALSO NOTED THAT THE ROOT CAUSE OF THE HEAT WAS FROM WORN OUT BEARINGS AND THE COMPONENT DAMAGE WAS CAUSED FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS HEATING UP. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 844959 | 7.5CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | THE ANSPACH EFFORT LLC | QD8-S | 00845384007159 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |