FDA Adverse Event Malfunction Summary report: N

SOVEREIGN SPINAL SYSTEM

MDR report key: 7062542 · Received November 28, 2017

Report

Report Number
1030489-2017-02396
Event Type
Malfunction
Date Received
November 28, 2017
Date of Event
November 1, 2017
Report Date
March 1, 2018
Manufacturer
MSD DEGGENDORF MFG
Product Code
OVD
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS: THE SOVEREIGN SPACER WAS RETURNED WITH A CRACK RUNNING FROM THE UPPER SIDE OF THE SCREW HOLE 180 DEGREES. THERE WERE WITNESS MARKS LOCATED UNDER THE SHELF OF THE SCREW HOLE. THIS INDICATES THAT THE SCREW WAS PLACED TOO DEEP EITHER FROM FORCE OR MISALIGNMENT. ONCE THE SCREW GETS UNDER THE SHELF IT WILL START TO PLACE PRESSURE OUTWARDLY ON THE IMPLANT LEADING TO THE IMPLANT CRACKING TO RELIEVE THE STRESS. PRODUCT IMAGE REVIEW: SUBMITTED IMAGES APPEAR TO DISPLAY CRACKED INTERBODY SPACER. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT CODE: OVD - INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR THIS PART IS NOT APPROVED FOR USE IN THE UNITED STATES; HOWEVER A LIKE DEVICE CATALOG # 7968212, 510K #K091813 AND UPN (B)(4) WAS APPROVED FOR SALE IN THE UNITED STATES. PRODUCT WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNSPECIFIED SURGERY AT L5-S1 DUE TO SPONDYLOLISTHESIS. INTRA-OPERATIVELY, THE CAGE CRACKED DURING FINAL TIGHTENING. THE CAGE WAS THEN REMOVED AND REPLACED WITH ANOTHER ONE OF THE SAME DIMENSION WITHOUT ANY PROBLEM. THERE WAS NO FRAGMENT OF THE DEVICE REMAINING INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
843861 SOVEREIGN SPINAL SYSTEM OVD MSD DEGGENDORF MFG NA 204755275

Patients

Seq Age Sex Outcome Treatment
1 51 YR