EMAX 2 PLUS MOTOR
Report
- Report Number
- 1045834-2017-50021
- Event Type
- Malfunction
- Date Received
- November 27, 2017
- Report Date
- October 30, 2017
- Manufacturer
- THE ANSPACH EFFORT LLC
- Product Code
- HBC
- UDI-DI
- 00845384002055
- PMA / PMN Number
- K080802
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
REPORTER'S FULL NAME WAS NOT PROVIDED. REPORTER¿S PHONE NUMBER: (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE DEVICE HAD WORN COMPONENTS - BEARINGS. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED FROM (B)(6) THAT THE MOTOR DEVICE DID NOT WORK AND HAD AN E6 (OVERHEAT WARNING) ERROR CODE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT IS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 838218 | EMAX 2 PLUS MOTOR | MOTOR, DRILL, ELECTRIC - HANDPIECE | HBC | THE ANSPACH EFFORT LLC | EMAX2PLUS | 00845384002055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |