FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ SYRINGE LUER-LOK¿

MDR report key: 7060318 · Received November 27, 2017

Report

Report Number
1213809-2017-00277
Event Type
Malfunction
Date Received
November 27, 2017
Date of Event
October 27, 2017
Report Date
June 11, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096580
PMA / PMN Number
K110771
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER PHONE#: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. DHR REVIEW FOR BATCH# 7060629 (P/N (B)(4)): MANUFACTURING DATES: 03/22/2017 TO 03/23/2017. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 7060629 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. BASED ON THE SEVERITY ASSIGNED TO THIS COMPLAINT AND THE RESULTS OF THE COMPLAINT LOT HISTORY CHECK, CAPA IS NOT REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS PRODUCT AND CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BD PLASTIPAK¿ SYRINGE LUER-LOK¿ WAS FOUND BEFORE USE WITH SCALE MARKING MISALIGNED, MAKING THIS IMPOSSIBLE TO MEASURE VOLUME. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
839022 BD PLASTIPAK¿ SYRINGE LUER-LOK¿ HYPODERMIC SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 7060629 30382903096580

Patients

Seq Age Sex Outcome Treatment
1 Other