FDA Adverse Event Malfunction Summary report: N

BD SYRINGE WITH NEEDLE

MDR report key: 7060108 · Received November 27, 2017

Report

Report Number
3002682307-2017-00099
Event Type
Malfunction
Date Received
November 27, 2017
Date of Event
October 31, 2017
Report Date
November 6, 2017
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SAMPLE EVALUATION - WE HAVE BEEN PROVIDED WITH THE AFFECTED SAMPLE. A LEAKAGE THROUGH THE PLUNGER ROD WAS OBSERVED IN THIS PROVIDED SAMPLE. WE COULD NOT DETERMINE THE EXACT ROOT CAUSE DUE TO THE PRESENCE OF DRY BLOOD INSIDE THE SYRINGE. THE SAMPLES WERE NOT DECONTAMINATED. BHR REVIEW WE HAVE REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QA NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2011 ((B)(6) 2017). SYRINGES WERE ASSEMBLED IN MACHINE, Nº4252, AND Nº4251, IN LOT #7089470 ((B)(6) 2017), IN LOT #7100073 ((B)(6) 2017). RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARRELS LOTS #7100035, AND #7089485 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #7114316, #7100040, #7089486, AND #7083123 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. ROOT CAUSE ANALYSIS: WE CONCLUDE THAT THE CAUSE OF THE PROBLEM COULD BE PRODUCED AS A CONSEQUENCE OF A DAMAGE IN THE PLUNGER LIP. THIS COULD BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR IN THE PLUNGER ASSEMBLY MACHINE. CONFIRMATION: THE PROVIDED SAMPLE PRESENTED THE REPORTED ISSUE. WE COULD CONFIRM THE REPORTED ISSUE. CAPA DETERMINATION: NO - BASED ON AN EVALUATION OF SEVERITY AND OCCURRENCE IT WAS DETERMINATE THAT NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NURSE FOUND LEAKAGE FROM THE PLUNGER ROD OF THE BD SYRINGE WITH NEEDLE. FOUND DURING USE. NO SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
838985 BD SYRINGE WITH NEEDLE HYPODERMIC SYRINGE AND NEEDLE FMF BECTON DICKINSON, S.A. 1704136

Patients

Seq Age Sex Outcome Treatment
1 Other