FDA Adverse Event
Injury
Summary report: N
COOK AIRWAY EXCHANGE CATHETERS EF
MDR report key: 705621
·
Received April 27, 2006
Report
- Report Number
- 1820334-2006-00086
- Event Type
- Injury
- Date Received
- April 27, 2006
- Date of Event
- March 22, 2006
- Report Date
- March 28, 2006
- Manufacturer
- COOK, INC.
- Product Code
- BTR
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
WHILE EXCHANGING A 41 ET TUBE FOR A 39 ET TUBE AT THE END OF A LEAF LOWER LOBECTOMY PROCEDURE, RESISTANCE WAS ENCOUNTERED. THE CAE WAS PULLED FIRMLY AND REMOVED WITH THE TUBE. A VISUAL INSPECTION OF THE DISTAL END OF THE CAE FOUND THE TIP APPEARED TO BE MISSING. UPON VISUALIZATION OF THE TRACHEAL LUMEN, USING A 4MM FLEXIBLE BRONCHOSCOPE, A PIECE OF THE CAE, APPROX 2.5CM LONG WAS SEEN PROXIMALLY TO THE BRONCHIAL LUMEN OF THE DLT. FURTER VISUALIZATION REVEALED THE PIECE HAD BEEN ROTATED 180 DEGREES. UTILIZING SUCTION, RIGID AND FLEXIBLE BRONCHOSCOPY AND VARIOUS GRASPERS AND FORCEPS, THE PIECE WAS REMOVED TWO HOURS LATER AND THE OPERATION PROCEEDED AS PLANNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COOK AIRWAY EXCHANGE CATHETERS EF | SPECIAL SET | BTR | COOK, INC. | NA | 1521788 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |