HEARTWARE VENTRICULAR ASSIST SYSTEM- CONTROLLER AC ADAPTER
Report
- Report Number
- 3007042319-2017-04238
- Event Type
- Malfunction
- Date Received
- November 22, 2017
- Date of Event
- September 2, 2015
- Report Date
- November 22, 2017
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT EVENT SUMMARY: ONE CONTROLLER AC (CAC) ADAPTER WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. THE REPORTED EVENT COULD NOT BE CONFIRMED; THE CAC ADAPTER OPERATED AS EXPECTED DURING BENCH TESTING. ANALYSIS OF THE DEVICE REVEALED THAT THE DEVICE MET SPECIFICATIONS; THE DEVICE PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. THE REPORTED EVENT COULD NOT BE DUPLICATED AT THE BENCH LEVEL. THE REPORTED EVENT COULD NOT BE CONFIRMED DURING TESTING; NO MECHANICAL ISSUES WERE FOUND DURING TESTING. THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF EVENTS, WHICH WAS IN RESPONSE TO AN UPDATE TO THE MDR DECISION CRITERIA. THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE CONTROLLER AC (CAC) ADAPTER DID NOT "CLICK" IN AND THAT IT DID NOT FIT WELL. THE CAC ADAPTER WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 833596 | HEARTWARE VENTRICULAR ASSIST SYSTEM- CONTROLLER AC ADAPTER | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1430DE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |