FDA Adverse Event Injury Summary report: N

VERSAFITCUP DOUBLE MOBILITY LINER Ø 52/28

MDR report key: 7052727 · Received November 22, 2017

Report

Report Number
3005180920-2017-00686
Event Type
Injury
Date Received
November 22, 2017
Date of Event
October 23, 2017
Report Date
November 22, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
UDI-DI
07630030807336
PMA / PMN Number
K083116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 17 NOVEMBER 2017 THE MEDICAL AFFAIRS PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: FEMORAL HEAD AND POLYETHYLENE DOUBLE MOBILITY LINER REVISION OCCURRED 7 YEARS AFTER PRIMARY TOTAL HIP ARTHROPLASTY. PRE-REVISION XRAYS ARE NOT SUPPLIED. ACCORDING TO THE REPORT, THE PATIENT WAS COMPLAINING ABOUT PAIN. OSTEOLYSIS MAY BE THE CONSEQUENCE OF POLYETHYLENE WEAR THAT IS COMMONLY SEEN AND DESCRIBED IN LITERATURE, HOWEVER WEAR WAS NOT MEASURED IN THIS CASE AND THEREFORE CORRELATION IS NOT ESTABLISHED. BATCH REVIEW PERFORMED ON 20 NOVEMBER 2017. LOT 092227: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25 SEPTEMBER 2009. EXPIRATION DATE: 2014-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ON 22 NOVEMBER 2017 THE R&D PROJECT MANAGER PERFORMED A PRELIMINARY INVESTIGATION BASED ON THE AVAILABLE IMAGES AND COMMENTED AS FOLLOWS: THE IMAGES SHOW THE LINER EXPLANTED: NO PARTICULAR SIGNS ARE VISIBLE, THE HEAD IS INTERNAL AND SOME BLOOD IS PRESENT IN THE EXTERNAL SURFACE AND INTERNALLY. FROM THE IMAGES IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF PAIN. THE PATIENT PRESENTED WITH EARLY STAGE OSTEOLYSIS AROUND THE PROXIMAL PORTION OF THE FEMUR. THE SURGEON REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
833004 VERSAFITCUP DOUBLE MOBILITY LINER Ø 52/28 DOUBLE MOBILITY LINER MEH MEDACTA INTERNATIONAL SA 092227 07630030807336

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention