FDA Adverse Event
Malfunction
Summary report: N
MONACO
MDR report key: 7052006
·
Received November 22, 2017
Report
- Report Number
- 1937649-2017-00007
- Event Type
- Malfunction
- Date Received
- November 22, 2017
- Report Date
- November 22, 2017
- Manufacturer
- ELEKTA INC
- Product Code
- MUJ
- UDI-DI
- 00858164002152
- PMA / PMN Number
- K151233
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. THE ISSUE WAS DETERMINED TO BE A DEFECT IN THE PRODUCT AND WILL BE FIXED IN A LATER VERSION.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT MONACO WILL DISPLAY THE INCORRECT MU FOR A 3D PLAN. BASED ON THE AVAILABLE INFORMATION THERE HAS BEEN NO MISTREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 831819 | MONACO | SYSTEM,PLANNING,RADIATION THERAPY TREATMENT | MUJ | ELEKTA INC | 00858164002152 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |