FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME

MDR report key: 7051160 · Received November 21, 2017

Report

Report Number
1037905-2017-00703
Event Type
Malfunction
Date Received
November 21, 2017
Date of Event
October 24, 2017
Report Date
December 19, 2017
Manufacturer
COOK ENDOSCOPY
Product Code
KNS
UDI-DI
00827002319035
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: ERBE ELECTROSURGICAL GENERATOR AND COOK FUSION ULTRA SHORT WIRE GUIDE (FSW-35) INVESTIGATION EVALUATION: AN EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORT ON INCORRECT CUTTING WIRE ORIENTATION. DURING OUR LABORATORY ANALYSIS, THE SPHINCTEROTOME WAS ADVANCED THROUGH A DUODENOSCOPE THAT IS PLACED IN A SIMULATED BILIARY POSITION. THE DUODENOSCOPE HAS AN ACCESSORY CHANNEL THAT IS 4.2 MM IN DIAMETER (MODEL NUMBER OLYMPUS TJF-160V). THE CATHETER EXITED THE ENDOSCOPE WITH THE CUTTING WIRE FACING 9 O'CLOCK. THE DEVICE WAS THEN BOWED AND THE CUTTING WIRE WAS FACING 3 O'CLOCK (APPROPRIATE ORIENTATION IS APPROXIMATELY 11:00 - 1:00 O'CLOCK). THE SPHINCTEROTOME CATHETER WAS SUBJECTED TO A CLOSE VISUAL EXAMINATION AND TWISTING OF THE TUBING WAS OBSERVED AT THE DISTAL END. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A NONCONFORMITY THAT COULD BE RELATED TO THE OBSERVATION REPORTED BY THE USER WAS FOUND. THE DEVICE GOES THROUGH SEVERAL INSPECTIONS IN MANUFACTURING, FINAL QUALITY CONTROL, AND PACKAGING DEPARTMENTS PRIOR TO LEAVING THE FACILITY IN AN EFFORT TO ENSURE PROPER ORIENTATION. THE INSPECTION PROCESSES WOULD HAVE REMOVED ANY PRODUCTS HAVING THIS NONCONFORMANCE PRIOR TO DISTRIBUTION. THEREFORE, IT IS UNKNOWN HOW OR AT WHAT POINT THE TUBING BECAME TWISTED AT THE DISTAL END. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION, OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: ¿NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF SPHINCTEROTOME IN AN ATTEMPT TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE.¿ OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER: "UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME. CAREFULLY REMOVE PRECURVED STYLET FROM CANNULATING TIP." THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: ERBE ELECTROSURGICAL GENERATOR AND COOK FUSION ULTRA SHORT WIRE GUIDE (FSW-35). INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A NONCONFORMITY THAT COULD BE RELATED TO THE OBSERVATION REPORTED BY THE USER WAS FOUND IN THE DHR FOR TWISTING. THE DEVICE GOES THROUGH SEVERAL INSPECTIONS IN MANUFACTURING, FINAL QUALITY CONTROL (FQC) AND PACKAGING DEPARTMENTS PRIOR TO LEAVING THE FACILITY IN AN EFFORT TO ENSURE PROPER ORIENTATION. THE INSPECTION PROCESSES WOULD HAVE REMOVED ANY PRODUCTS HAVING THIS NONCONFORMANCE PRIOR TO DISTRIBUTION. THEREFORE, IT IS UNKNOWN HOW OR AT WHAT POINT THE TUBING BECAME TWISTED AT THE DISTAL END. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: ¿NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF SPHINCTEROTOME IN AN ATTEMPT TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE.¿ OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER: "UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME. CAREFULLY REMOVE PRECURVED STYLET FROM CANNULATING TIP." THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME SPHINCTEROTOME (FS-OMNI). THEY NOTICED THE FS-OMNI WAS DISTORTED AND IN THE INCORRECT/ OPPOSITE ORIENTATION ONCE THE DEVICE WAS ADVANCED THROUGH THE WORKING CHANNEL OF THE ENDOSCOPE. THE FS-OMNI WAS THEN REMOVED TO BE SENT BACK. THE PROCEDURE WAS COMPLETED USING ANOTHER FS-OMNI SPHINCTEROTOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
827649 FUSION OMNI-TOME KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS COOK ENDOSCOPY 00827002319035

Patients

Seq Age Sex Outcome Treatment
1