8CM ANGLE ATTACHMENT
Report
- Report Number
- 1045834-2017-50014
- Event Type
- Malfunction
- Date Received
- November 21, 2017
- Report Date
- October 27, 2017
- Manufacturer
- THE ANSPACH EFFORT LLC
- Product Code
- HBC
- UDI-DI
- 00845384006626
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE REPORTER'S FULL NAME WAS NOT PROVIDED. REPORTER'S PHONE NUMBER: (B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DATE DEVICE RETURNED TO MANUFACTURER: THE DEVICE RETURNED DATE WAS INADVERTENTLY OMITTED FROM THE INITIAL REPORT. THE DATE HAS BEEN UPDATED AS 10/27/2017. THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THE BEARINGS WERE WORN. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED FROM (B)(6) THAT THE MOTOR DEVICE GENERATED HEAT (55 DEGREES CELSIUS /3 MIN). IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 828747 | 8CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | THE ANSPACH EFFORT LLC | QD8 | 00845384006626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |