TRUETRACK
Report
- Report Number
- 1000113657-2017-01981
- Event Type
- Malfunction
- Date Received
- November 21, 2017
- Date of Event
- October 31, 2017
- Report Date
- November 21, 2017
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292001797
- PMA / PMN Number
- K032657
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). (B)(4).
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 172, 182, 223, 153 AND 167 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 107 - 120 MG/DL. CUSTOMER STATED THE HE HAS BEEN OBTAINING THE HIGH RESULTS SINCE HE STARTED USING THE TEST STRIPS TWO - THREE WEEKS AGO. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. CUSTOMER STATED THAT HE HAD GONE TO THE DOCTOR FOR HIS NORMAL VISIT AND TOLD THE DOCTOR ABOUT THE METER. THE DOCTOR HAD DONE BLOOD WORK; CUSTOMER COULD NOT REMEMBER WHAT THE RESULTS WERE BUT STATED THAT THE DOCTOR TOOK HIM OFF HIS HIGH DOSAGE OF METFORMIN SINCE HIS BLOOD WORK WAS GOOD. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 06/21/2019 AND OPEN VIAL DATE AT TIME OF CALL IS TWO - THREE WEEKS. CUSTOMER HAD THE CORRECT CODE CHIP IN THE METER. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (TIME NOT SET): RESULT 1: 172 MG/DL DATE: (B)(6) 2017TIME:1:18PM FASTING, RESULT 2 :182 MG/DL DATE: (B)(6) 2017TIME: 1:49PM FASTING, RESULT 3 :223 MG/DL DATE: (B)(6) 2017TIME:1:48PM FASTING, RESULT 4 :153 MG/DL DATE:(B)(6) 2017 TIME: 1:24PM FASTING, RESULT 5 :167 MG/DL DATE:(B)(6) 2017 TIME: 1:09PM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 830952 | TRUETRACK | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUETRACK | RU5065 | 00021292001797 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |