FDA Adverse Event Malfunction Summary report: N

6.5CM ADULT CRANI ATTACHMT

MDR report key: 7048114 · Received November 21, 2017

Report

Report Number
1045834-2017-50011
Event Type
Malfunction
Date Received
November 21, 2017
Report Date
October 30, 2017
Manufacturer
THE ANSPACH EFFORT LLC
Product Code
HBC
UDI-DI
845384001720
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTER'S FULL NAME WAS NOT PROVIDED. REPORTER'S PHONE NUMBER: (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THE BEARINGS WERE WORN. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

CORRECTION: DEVICE AVAILABILITY: THE DEVICE AVAILABILITY WAS INADVERTENTLY DOCUMENTED AS 10/31/2017 IN THE INITIAL REPORT. THE DATE RETURNED TO MANUFACTURER WAS CORRECTED TO 11/6/2017. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE INSIDE PARTS OF THE CRANIOTOME DEVICE DROPPED OFF. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
829922 6.5CM ADULT CRANI ATTACHMT MOTOR, DRILL, ELECTRIC - CRANIOTOME HBC THE ANSPACH EFFORT LLC 845384001720

Patients

Seq Age Sex Outcome Treatment
1