FDA Adverse Event Death Summary report: N

REFOBACIN BONE CEMENT R 1X40G

MDR report key: 7046628 · Received November 20, 2017

Report

Report Number
3006946279-2017-00250
Event Type
Death
Date Received
November 20, 2017
Date of Event
December 29, 2016
Report Date
February 9, 2018
Manufacturer
BIOMET FRANCE S.A.R.L.
Product Code
MBB
PMA / PMN Number
PK150850
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(4). REPORT SOURCE- FOREIGN. THE EVENT OCCURRED IN (B)(4). THE PRODUCT WITHIN THIS REPORT IS A COMBINATION PRODUCT. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. RESERVE SAMPLE WAS EVALUATED UNDER STANDARDIZED CONDITIONS AND NO FAILURE OF THE BONE CEMENT WAS FOUND; THE PRODUCT FUNCTIONED AS INTENDED. INVESTIGATION RESULTS CONCLUDED THAT THE REPORTED EVENT WAS LIKELY DUE TO A KNOWN RISK, KNOWN AS IMPLANTATION SYNDROME. THIS RISK IS ADDRESSED IN THE INSTRUCTIONS FOR USE (IFU) OF THIS DEVICE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION HAS BEEN PERFORMED, CONSISTING OF A PRODUCT ANALYSIS AND A DOCUMENTARY REVIEW. THE DOCUMENTARY REVIEW SHOWED THAT PRODUCTS WERE MANUFACTURED ACCORDING TO THE PRE-DEFINED SPECIFICATIONS OF BIOMET (B)(4). PRODUCT ANALYSIS ON RETAINED SAMPLES FROM THE SAME BATCH SHOWED NO UNUSUAL BEHAVIOUR DURING MIXING, HANDLING OR SETTING. THE MOST PROBABLE REASON FOR DEATH IS THE CEMENT IMPLANTATION SYNDROME. THIS RISK IS DESCRIBED IN THE IFU PROVIDED WITH THE PRODUCT. NO NON-CONFORMITY HAS BEEN DETECTED CONCERNING THE PRODUCTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, THE PATIENT EXPERIENCED DECREASE OF THE BLOOD PRESSURE FOLLOWED BY A CARDIAC ARREST AS THE PROSTHESIS WAS CEMENTED. THE PATIENT WAS RECOVERED AFTER CARDIOVASCULAR REANIMATION. UPON WAKING, THE PATIENT PRESENTED WITH SEVERAL CARDIAC ARRESTS THAT GAVE RISE TO SEVERAL REANIMATIONS DURING 2HOURS AND 5 MINUTES. AFTER CONSULTATION, DECISION TO STOP THE REANIMATION WAS MADE. THE PATIENT EXPIRED. NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
825062 REFOBACIN BONE CEMENT R 1X40G BONE CEMENT, ANTIBIOTIC MBB BIOMET FRANCE S.A.R.L. N/A A622CA1904

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death REFOBACIN BONE CEMENT R 1X40G LOT 3003940001