FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 7042726 · Received November 17, 2017

Report

Report Number
1045834-2017-50005
Event Type
Malfunction
Date Received
November 17, 2017
Report Date
October 27, 2017
Manufacturer
THE ANSPACH EFFORT LLC
Product Code
HBC
UDI-DI
00845384002055
PMA / PMN Number
K080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THE LOCKING COMPONENTS WERE DAMAGED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO PREMATURE WEAR FROM NORMAL USE SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

THE REPORTER'S FULL NAME WAS NOT PROVIDED. REPORTER'S PHONE NUMBER:(B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE MOTOR DEVICE GENERATED HEAT AND HAD A ROTATION SHORTAGE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
819836 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC THE ANSPACH EFFORT LLC EMAX2PLUS N/A 00845384002055

Patients

Seq Age Sex Outcome Treatment
1