FDA Adverse Event Malfunction Summary report: N

SELF CONSTRAINED AWL (STAINLESS STEEL)

MDR report key: 7039007 · Received November 16, 2017

Report

Report Number
2027467-2017-00083
Event Type
Malfunction
Date Received
November 16, 2017
Date of Event
October 23, 2017
Report Date
October 24, 2017
Manufacturer
ALPHATEC SPINE INC
Product Code
HWJ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE RETURNED DEVICE REVEALED A FRAGMENT OF THE DISTAL TIP HAD BROKEN AND DETACHED FROM THE INSTRUMENT. THE REMAINDER OF THE POINTED EDGES ARE SLIGHTLY BENT/DEFORMED AND NO LONGER SHARP. THE MANIPULATION APPEARS TO BE CONSISTENT WITH THAT OF POSSIBLY IMPACTING/STRIKING WITH EXCESSIVE FORCE AT THE PROXIMAL END OF THE INSTRUMENT. THE SELF CONSTRAINED AWL IS DESIGNED TO CREATE A PILOT HOLE THROUGH CORTICAL BONE BY PUSHING DOWN ON THE HANDLE OF THE INSTRUMENT.

Description of Event or Problem · 1

AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THAT THE TIP OF THE AWL IS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
815607 SELF CONSTRAINED AWL (STAINLESS STEEL) HWJ HWJ ALPHATEC SPINE INC 61730 5873701

Patients

Seq Age Sex Outcome Treatment
1 Other