FDA Adverse Event
Malfunction
Summary report: N
SELF CONSTRAINED AWL (STAINLESS STEEL)
MDR report key: 7039007
·
Received November 16, 2017
Report
- Report Number
- 2027467-2017-00083
- Event Type
- Malfunction
- Date Received
- November 16, 2017
- Date of Event
- October 23, 2017
- Report Date
- October 24, 2017
- Manufacturer
- ALPHATEC SPINE INC
- Product Code
- HWJ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
AN EVALUATION OF THE RETURNED DEVICE REVEALED A FRAGMENT OF THE DISTAL TIP HAD BROKEN AND DETACHED FROM THE INSTRUMENT. THE REMAINDER OF THE POINTED EDGES ARE SLIGHTLY BENT/DEFORMED AND NO LONGER SHARP. THE MANIPULATION APPEARS TO BE CONSISTENT WITH THAT OF POSSIBLY IMPACTING/STRIKING WITH EXCESSIVE FORCE AT THE PROXIMAL END OF THE INSTRUMENT. THE SELF CONSTRAINED AWL IS DESIGNED TO CREATE A PILOT HOLE THROUGH CORTICAL BONE BY PUSHING DOWN ON THE HANDLE OF THE INSTRUMENT.
Description of Event or Problem · 1
AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THAT THE TIP OF THE AWL IS BROKEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 815607 | SELF CONSTRAINED AWL (STAINLESS STEEL) | HWJ | HWJ | ALPHATEC SPINE INC | 61730 | 5873701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |