OAE PEDIATRIC FOAM EAR TIPS
Report
- Report Number
- 3018859-2017-00916
- Event Type
- Malfunction
- Date Received
- November 16, 2017
- Report Date
- October 17, 2017
- Manufacturer
- NATUS MEDICAL INC.
- Product Code
- GWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THIS ISSUE WAS INVESTIGATED THROUGH CORRECTIVE ACTIONS. THE ROOT CAUSE OF THIS ISSUE WAS CONFIRMED TO BE INADEQUATE APPLICATION OF THE ADHESIVE WHICH SECURES THE FOAM TO THE LUMEN TUBE.
THE CUSTOMER WAS PROVIDED REPLACEMENT FOAM TIPS. NATUS INSTRUCTIONS INCLUDE SELECTING THE APPROPRIATE SIZE EAR TIP FOR THE PATIENT, AS WELL AS INSTRUCTIONS REGARDING USING LARGER SIZE FOAM TIPS FOR INSERTION INTO THE EAR CANAL. NATUS IS INVESTIGATING THIS ISSUE THROUGH CORRECTIVE ACTIONS AT THIS TIME.
THE CUSTOMER REPORTED TO NATUS MEDICAL THAT THE FOAM TIPS OF THE PEDIATRIC FOAM EAR TIPS ARE SEPARATING FROM THE PLASTIC TUBING. THERE HAS BEEN NO REPORT OF PATIENT DEATH, INJURY OR DELAY IN PATIENT TREATMENT. NO ENVIRONMENTAL OR SAFETY CONCERNS HAVE ALSO BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 817227 | OAE PEDIATRIC FOAM EAR TIPS | OAE PEDIATRIC FOAM EAR TIPS | GWJ | NATUS MEDICAL INC. | 203211 | N070116-11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |