Description of Event or Problem · 1
INSTRUMENT MALFUNCTION RESULTING IN REPORTING OF ERRONEOUS TEST RESULTS -CK MB AND TROPONIN I-. POST EVENT, ABBOTT TECHNICAL SERVICE DETERMINED THE MALFUNCTION TO BE INSUFFICIENT DISPENSING OF SOLUTION #3 -MATRIX CELL WASH-; HOWEVER, DURING THE PERIOD OF INSTRUMENT MALFUNCTION, RESULTS WERE PRODUCED IN 2 SEPARATE TEST RUNS WHICH CONTAINED NO ERROR CODES OR WARNING FLAGS INDICATIVE OF INSTRUMENT MALFUNCTION TO ALERT THE TECH THAT RESULTS WERE QUESTIONABLE. AS PER STANDARD OPERATING PROCEDURE, CRITICAL TESTS WERE REPEATED AND VERIFIED. DISCOVERY OF MALFUNCTION COMMENCED WHEN THE MEDICAL TECHNOLOGIST RECOGNIZED AN UNUSUAL TREND IN THE NUMBER OF POSITIVE TROPONINS, CK MBS, AND HCGS, QUALITY CONTROL WAS RUN WHICH SUBSEQUENTLY FAILED. SAMPLE WAS RETESTED ON 2ND AXSYM WHICH YIELDED CORRECTED RESULT. CORRECTED REPORT WAS ISSUED AND PROVIDER WAS NOTIFIED. ABBOTT DIAGNOSTICS WERE ENGAGED FOR PROBLEM RESOLTUION. FULL PREVENTIVE MAINTENANCE WAS PERFORMED BY ABBOTT IN 04/06. REGIONAL LABORATORY MGR REQUESTED PLACEMENT OF SENSORS ON ASPIRATING STRAW -REAGENT 3- ALONG WITH DEFINITIVE REPORT WITH REGARD TO PRODUCT IMPROVEMENT FROM ABBOTT'S ENGINEERING DIVISION.