FDA Adverse Event
Malfunction
Summary report: N
INTELLIVUE MP50
MDR report key: 7027829
·
Received November 14, 2017
Report
- Report Number
- 7027829
- Event Type
- Malfunction
- Date Received
- November 14, 2017
- Date of Event
- July 28, 2017
- Report Date
- September 29, 2017
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- BZQ
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
BIO MED LOG: ELECTRONIC HEALTH RECORD (EHR) NOT DOWNLOADING TO MONITOR. MONITOR TESTED, CONFIRMED DATA NOT TRANSFERRING FROM MONITOR TO EHR. PHILIPS TECH SUPPORT CALLED, PARTS ORDERED AND INSTALLED. SIX WEEKS LATER - ABOVE ISSUE STILL EXISTS. MORE TROUBLE SHOOTING WITH PHILIPS SUPPORT AND EHR IT BUT NO SOLUTION. THREE MONTHS PASSED AND ISSUE CONTINUES. NEW NETWORK CARD OBTAINED AND INSTALLED, REPLACING CARD PURCHASED 6 MONTHS PRIOR. ISSUE DID NOT RESOLVE. FINALLY AFTER 4 MONTHS, DECISION MADE TO REPLACE THE MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 808558 | INTELLIVUE MP50 | BEDSIDE PHYSIOLOGIC MODULAR MONITOR | BZQ | PHILIPS MEDICAL SYSTEMS | M8004A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |