Description of Event or Problem · 1

BIO MED LOG: ELECTRONIC HEALTH RECORD (EHR) NOT DOWNLOADING TO MONITOR. MONITOR TESTED, CONFIRMED DATA NOT TRANSFERRING FROM MONITOR TO EHR. PHILIPS TECH SUPPORT CALLED, PARTS ORDERED AND INSTALLED. SIX WEEKS LATER - ABOVE ISSUE STILL EXISTS. MORE TROUBLE SHOOTING WITH PHILIPS SUPPORT AND EHR IT BUT NO SOLUTION. THREE MONTHS PASSED AND ISSUE CONTINUES. NEW NETWORK CARD OBTAINED AND INSTALLED, REPLACING CARD PURCHASED 6 MONTHS PRIOR. ISSUE DID NOT RESOLVE. FINALLY AFTER 4 MONTHS, DECISION MADE TO REPLACE THE MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
808558 INTELLIVUE MP50 BEDSIDE PHYSIOLOGIC MODULAR MONITOR BZQ PHILIPS MEDICAL SYSTEMS M8004A

Patients

Seq Age Sex Outcome Treatment
1