FDA Adverse Event Injury Summary report: N

GAP ENDO-EXO MEDULLARY SYSTEM

MDR report key: 7024387 · Received November 13, 2017

Report

Report Number
3000327445-2017-00005
Event Type
Injury
Date Received
November 13, 2017
Date of Event
October 14, 2017
Report Date
November 10, 2017
Manufacturer
PEGA MEDICAL, INC.
Product Code
HSB
PMA / PMN Number
K160545
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS GAP-N56-28 NAIL WAS INSERTED INTO A (B)(6) YEAR-OLD OVERWEIGHT FEMALE PATIENT WITH OSTEOPOROTIC BONE (WEAK BONE). THE FEMUR PRESENTED WITH A NARROW SHAFT AND A VALGUS DEFORMATION. THE FAILURE OCCURRED APPROXIMATELY THREE WEEKS POST-SURGERY WHILE THE PATIENT WAS SITTING TO PUT HER SHOE ON. THE WEIGHT OF THE PATIENT (AROUND 60 KG) WAS WELL ABOVE THE WEIGHT LIMIT OF THE NAIL, WHICH IS AROUND 40 KG. FURTHERMORE, THE NAIL WAS INSERTED IN RETROGRADE INSTEAD OF ANTEGRADE. ANTEROGRADE INSERTION IS THE PREFERRED APPROACH IN PATIENTS WITH PROXIMAL FRACTURES SUCH AS THIS CASE SINCE IT MAXIMIZES THE RESISTANCE PROXIMALLY WHERE IT IS MOST NEEDED. RETROGRADE INSERTION PLACES THE WEAKEST SECTION OF THE NAIL (THE TIP OF THE NAIL) PROXIMALLY WHERE THE HIGHEST LOADS ARE EXPERIENCED DUE TO THE PROXIMITY OF THE FRACTURE OR OSTEOTOMY AND THE LEVER ARM OF THE HIP. THE SURGEON HAD NO CHOICE BUT TO INSERT THE NAIL IN RETROGRADE DUE TO SPECIFIC PATIENT CONDITIONS (DISTAL OSTEOTOMY). THE COMBINATION OF HIGH LOADS DUE TO THE PATIENT BEING OVERWEIGHT AND THE RETROGRADE INSERTION, CAUSED THE NAIL TO FAIL PROXIMALLY AT THE DISTAL LOCKING SCREW HOLE. THE BROKEN NAIL HAS BEEN SUCCESSFULLY RETRIEVED AND REPLACED FOR ANOTHER NAIL WITH A LARGER DIAMETER.

Description of Event or Problem · 1

A GAP-N56-28 NAIL WAS USED IN A (B)(6) YEAR-OLD OVERWEIGHT PATIENT TO CORRECT A VALGUS DEFORMITY. THE NAIL WAS INSERTED IN RETROGRADE FOLLOWING OSTEOTOMY AND CORRECTION OF THE RIGHT FEMUR. THREE WEEKS POST-SURGERY THE NAIL FAILED PROXIMALLY, SPECIFICALLY AT THE DISTAL SCREW HOLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
803610 GAP ENDO-EXO MEDULLARY SYSTEM INTRAMEDULLARY NAIL HSB PEGA MEDICAL, INC. 121218-010

Patients

Seq Age Sex Outcome Treatment
1 32 YR Required Intervention