HIP END EFFECTOR, VARIABLE ANGLE
Report
- Report Number
- 3005985723-2017-00540
- Event Type
- Malfunction
- Date Received
- November 10, 2017
- Date of Event
- November 7, 2017
- Report Date
- January 18, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K170593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: HIP END EFFECTOR COLLAR BROKE DURING IMPACTION. THA PROCEDURE. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION VISUAL INSPECTION CONFIRMS FAILED LOCKING MECHANISM WITH BALL BEARINGS FAILING. DISCOLORATION AND WEAR ARE SEEN ON THE 202862 HIP BALL RETAINER AND 202870 HIP CHUCK SLIDE ASSEMBLY. VISUAL INSPECTION CONFIRMS A SHEARED OFF 202866 HIP RELEASE KNOB. SEE ATTACHMENTS FOR IMAGES OF THE INSTRUMENT. DIMENSIONAL INSPECTION DIMENSIONAL INSPECTION WAS NOT COMPLETED AS VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION FUNCTIONAL INSPECTION AS NOT COMPLETED AS VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE 45 DEVICES WERE MANUFACTURED UNDER LOT NO 19430816 AND (B)(4) INCLUDING S/N (B)(4) ACCEPTED INTO FINAL STOCK ON 04/20/2016. A REVIEW OF (B)(4) REVEALED THAT THE NON-CONFORMANCE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED P/N 206967, LOT 19490816 SHOWS 1 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. (B)(4). CONCLUSIONS: ALLEGED FAILURE CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA 1450905 WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.
HIP END EFFECTOR COLLAR BROKE DURING IMPACTION. THA PROCEDURE.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
HIP END EFFECTOR COLLAR BROKE DURING IMPACTION. THA PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 800083 | HIP END EFFECTOR, VARIABLE ANGLE | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |