FDA Adverse Event Malfunction Summary report: N

BD ANGIOCATH¿ IV CATHETER 20G X 1.88IN

MDR report key: 7014800 · Received November 9, 2017

Report

Report Number
9610048-2017-00060
Event Type
Malfunction
Date Received
November 9, 2017
Date of Event
October 3, 2017
Report Date
December 5, 2017
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FOZ
UDI-DI
30382903811374
PMA / PMN Number
K950301
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

NO SAMPLES OR PHOTOS WERE RECEIVED FOR ANALYSIS OF THE INCIDENT IN QUESTION. THE FINAL ASSEMBLY BATCHES: 6146449SH, 6146457SH USED IN THE CLAIMED FINAL PRODUCT BATCHES: 6193418 OF ANGIOCATH 20G X 1.88 WW WERE CHECKED AND TESTED FOR PRESENCE OF "FOREIGN/NO-FOREIGN MATTER" AND THERE WERE NO RECORDS OF THIS DEFECT. THERE ARE NO QUALITY NOTIFICATION (QN) OR NON-CONFORMITY REPORT RECORDS OF FOREIGN MATTER, FOR THE LOTS INVOLVED IN THIS COMPLAINT, ACCORDING TO THE DHR OF THE LOT ABOVE. NOT CONFIRMED: BD WAS UNABLE TO CONFIRM THE INCIDENT IN QUESTION. AS NO PHOTOS OR SAMPLES RETURNED FROM THE CLIENT WERE RECEIVED FOR ANALYSIS, THE PRESENCE OF FOREIGN MATTER IN THE CATHETER COULD NOT BE CONFIRMED. BASED ON A SEVERITY ASSESSMENT AND OCCURRENCE IT WAS DETERMINED THAT NO CAPA IS REQUIRED AT THIS TIME. THE COMPLAINT WAS ADDED TO THE COMPLAINT DATABASE FOR TREND ANALYSIS, WHICH IS REGULARLY MONITORED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND DEPOSITED ON THE SURFACE OF SEVERAL BD ANGIOCATH¿ IV CATHETER 20G X 1.88IN PRIOR TO USE. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
794937 BD ANGIOCATH¿ IV CATHETER 20G X 1.88IN INTRAVASCULAR CATHETER FOZ BECTON DICKINSON AND CO. 6193418 30382903811374

Patients

Seq Age Sex Outcome Treatment
1 Other