FDA Adverse Event
Other
Summary report: N
*
MDR report key: 701350
·
Received March 24, 2006
Report
- Report Number
- 701350
- Event Type
- Other
- Date Received
- March 24, 2006
- Date of Event
- March 23, 2006
- Report Date
- March 24, 2006
- Manufacturer
- KARL STORZ ENDOSCOPY AMERICA, INC.
- Product Code
- KNW
- Report Source
- User Facility report
- Reporter Location
- OH, US
Narratives
Description of Event or Problem · 1
A PIECE OF THE BIOPSY FORCEPS LODGED IN PATIENT'S BRAIN DURING SURGERY. PIECE MAY BE IN BRAIN OR SOFT TISSUE. INJURY TO PATIENT, UNKNOWN, PATIENT TO RECEIVE POST OP MRI. UPON INSPECTION THE SCREW TO ATTACH THE OPENING MECHANISM TO THE FORCEP TIP IS MISSING. FROM THE EXAMINATION IT LOOKS LIKE THE OPENING MECHANISM MAY HAVE BEEN SLIGHTLY BENT CAUSING THE SCREW TO COME OUT. THIS DOES NOT APPEAR TO BE A MANUFACTURING FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | * | BIOPSY FORCEPS | KNW | KARL STORZ ENDOSCOPY AMERICA, INC. | 28162 PL | FG 01 RB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |