FDA Adverse Event Malfunction Summary report: N

VECTRA GRAFT, 6MM W/GRAD REINFORCE

MDR report key: 7010690 · Received November 8, 2017

Report

Report Number
2916596-2017-02677
Event Type
Malfunction
Date Received
November 8, 2017
Date of Event
October 17, 2017
Report Date
January 4, 2018
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PART AND LOT NUMBERS: PART # 10002-6050-001, LOT # 198404-1, QUANTITY: 23, LOT # 198404-2, QUANTITY: 2 . THE PRODUCTS WERE RETURNED FOR INVESTIGATION. THE EVALUATION IS NOT YET COMPLETE. THE EVENT OCCURRED IN (B)(6). NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION. THE PART NUMBER OF THE DEVICE WAS CORRECTED FROM 10002-6050-001 TO 10007188. THE REPORTED EVENT OF INCORRECT PACKAGE LABELING WAS CONFIRMED. THE CALL-OUT FOR THE DIAMETER AND LENGTH ON THE BOTTOM FLAP WAS INCORRECTLY LABELED AS 5.0 MM IN DIAMETER AND 40 CM IN LENGTH INSTEAD OF 6.0 MM IN DIAMETER AND 50 CM IN LENGTH. THE OTHER DIAMETER AND LENGTH CALL-OUTS ON THE LABELS WERE CORRECT. THE LABEL PRINTING FILE FOR THE PART'S LABEL WAS FOUND TO BE INCORRECT AND WAS REVISED TO CORRECTLY HAVE THE PROPER DIAMETER AND LENGTH ON THE BOTTOM FLAP. THE PRODUCTS WERE NOT IN USE AT THE TIME OF THE EVENT. THE ISSUE WITH THE OUTER CARTON LABELING WAS OBSERVED WHEN THE VECTRA GRAFTS WERE RECEIVED BY THE HOSPITAL. THE RETURNED PRODUCTS WERE REWORKED AND SENT BACK TO THE HOSPITAL. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT UPON RECEIPT OF THE PRODUCTS, INCORRECT LABELING WAS DISCOVERED ON THE OUTSIDE OF THE BOXES. THE DIAMETER AND LENGTH MEASUREMENTS WERE LISTED INCORRECTLY ON TWO SECTIONS OF THE LABEL ON THE OUTSIDE OF THE BOX. THE PATIENT LABELS AND LABELS ON THE TRAYS WERE CORRECT. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
791332 VECTRA GRAFT, 6MM W/GRAD REINFORCE VECTRA GRAFT DSQ THORATEC CORPORATION 198404-1

Patients

Seq Age Sex Outcome Treatment
1