FDA Adverse Event
Malfunction
Summary report: N
PICCOLO COMPOSITE HUMERAL NAIL
MDR report key: 7005739
·
Received November 7, 2017
Report
- Report Number
- 9615128-2017-00048
- Event Type
- Malfunction
- Date Received
- November 7, 2017
- Date of Event
- September 25, 2017
- Report Date
- November 7, 2017
- Manufacturer
- CARBOFIX ORTHOPEDICS LTD.
- Product Code
- HSB
- PMA / PMN Number
- K091425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCTION RECORDS OF THE NAIL DID NOT REVEAL DEVIATION FROM SPECIFICATION. THE INVOLVED NAIL WAS RECEIVED AND UNDERWENT EXAMINATION BY THE MANUFACTURER. ACCORDING TO THE FINDINGS, IT SEEMS THAT DURING NAIL INSERTION, EXCESSIVE FORCE WAS APPLIED ON THE PROXIMAL PART OF THE NAIL. IT SHOULD BE NOTED THAT IN CASES OF A DIFFICULT INSERTION, REAMING MAY BE PERFORMED TO FACILITATE THE INSERTION PROCESS.
Description of Event or Problem · 1
DURING ATTEMPT TO INSERT A PICCOLO COMPOSITE HUMERAL NAIL, THE THREAD AREA IN THE PROXIMAL PART OF THE NAIL (WHICH IS ATTACHED TO THE INSERTION HANDLE) WAS DAMAGED. THE NAIL WAS REMOVED AND REPLACED BY A LONGER CARBOFIX NAIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 787046 | PICCOLO COMPOSITE HUMERAL NAIL | INTRAMEDULLARY FIXATION ROD SYSTEM | HSB | CARBOFIX ORTHOPEDICS LTD. | QTN04157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |