FDA Adverse Event Malfunction Summary report: N

OPEN IMPLANT CLOSURE TOP

MDR report key: 7003349 · Received November 6, 2017

Report

Report Number
3012447612-2017-00617
Event Type
Malfunction
Date Received
November 6, 2017
Date of Event
October 6, 2017
Report Date
April 27, 2018
Manufacturer
ZIMMER BIOMET SPINE INC.
Product Code
NKB
PMA / PMN Number
PK131980
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: THE RETURNED CLOSURE TOP WAS EVALUATED. THERE WERE NO INDICATIONS OF DAMAGE ON THE DEVICE. THE CLOSURE TOP WAS FOUND TO WORK AS EXPECTED DURING A FUNCTIONAL ANALYSIS WITH MATING PARTS. THERE WAS NO FAILURE DETECTED. A REVIEW OF THE MANUFACTURING RECORDS DID NOT IDENTIFY ANY ISSUES WHICH WOULD HAVE CONTRIBUTED TO THIS EVENT.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION LED TO CORRECTIONS: LOT NUMBER AND UDI NUMBER. WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT.

Additional Manufacturer Narrative · 1

WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT.

Description of Event or Problem · 1

A CLOSURE TOP WAS FOUND TO NOT THREAD SMOOTHLY WITH A MATING SCREW. THIS WAS FOUND DURING A ROUTINE INSPECTION AND DID NOT HAVE ANY SURGICAL OR PATIENT IMPACTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
781827 OPEN IMPLANT CLOSURE TOP SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM NKB ZIMMER BIOMET SPINE INC. NA 87CS

Patients

Seq Age Sex Outcome Treatment
1