FDA Adverse Event Death Summary report: N

MOUNTAINEER 4.5MM UNIVERSAL JOINT TAP

MDR report key: 700267 · Received April 14, 2006

Report

Report Number
1526439-2006-00095
Event Type
Death
Date Received
April 14, 2006
Date of Event
May 24, 2005
Manufacturer
DEPUY SPINE, INC.
Product Code
LXH
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE SURGEON REPORTED THAT DUE TO HIGH ANGLE OF OCCIPUT AND ANATOMICAL LIMITATION THE SURGOEN USED THE UNIVERSAL JOINT TAP WITHOUT THE DRILL/TAP GUIDE WITH STOP MECHANISM. THE TAP PERFORATED THE BICORTICAL SKULL BASE DURING SUE. THE SCREW WAS PLACED, AS WAS THE REMAINING INSTRUMENTATION. THE PT WAS MONITORED. UPON REEXAMINATION THE PT HAD POST-OP INTRACRANIAL HEMATOMA/SWELLING AND LATER EXPIRED. THE EVENT IS NOT CONSIDERED TO BE DEVICE RELATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MOUNTAINEER 4.5MM UNIVERSAL JOINT TAP MANUAL SURGICAL INSTRUMENT LXH DEPUY SPINE, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death