FDA Adverse Event Malfunction Summary report: N

27 G X 1/2 IN. BD PRECISIONGLIDE¿ NEEDLE

MDR report key: 7002296 · Received November 5, 2017

Report

Report Number
1911916-2017-00295
Event Type
Malfunction
Date Received
November 5, 2017
Date of Event
October 17, 2017
Report Date
October 19, 2017
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051091
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER: ADDRESS UNAVAILABLE. BD CORPORATE ADDRESS USED. INVESTIGATION: INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. THE INVESTIGATION CONCLUDED BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS INVESTIGATION CONCLUSION: NO SAMPLES OR PHOTOS WERE PROVIDED. AN INVESTIGATION CANNOT BE COMPLETED. NO NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCH. DHR REVIEW: ASSEMBLY BATCH 6090856 HAD 50 VISUAL INSPECTIONS PERFORMED ON 2,650 PARTS WITH ZERO DEFECTS NOTED. SIX LEAK TESTS WERE PERFORMED ON 30 PARTS WITH ZERO FAILURES RECORDED. ROOT CAUSE DESCRIPTION: UNABLE TO DETERMINE WITHOUT SAMPLES.

Description of Event or Problem · 1

IT WAS REPORTED THAT TWO 27 G X 1/2 IN. BD PRECISIONGLIDE¿ NEEDLES LEAKED BEFORE USE. NO INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
781647 27 G X 1/2 IN. BD PRECISIONGLIDE¿ NEEDLE HYPODERMIC NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 6239515 30382903051091

Patients

Seq Age Sex Outcome Treatment
1 Other