FDA Adverse Event Injury Summary report: N

MEDSTREAM REFILL KIT

MDR report key: 7001258 · Received November 3, 2017

Report

Report Number
8031062-2017-00011
Event Type
Injury
Date Received
November 3, 2017
Report Date
September 5, 2017
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
LKK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS MDR, MFR REPORT #8031062-2017-00011, IS BEING SUBMITTED AS A FOLLOW-UP/FINAL REPORT TO INITIAL MDR, MFR REPORT #1226348 - 2017 - 10707. MFR REPORT #1226348 - 2017 - 10712 WAS SUBMITTED BY CODMAN NEUROSURGICAL ON SEPTEMBER 29, 2017. AS CODMAN NEUROSURGICAL WAS DIVESTED TO INTEGRA LIFE SCIENCES ON OCTOBER 01, 2017, CODMAN NEUROVASCULAR HAS TAKEN ON RESPONSIBILITY FOR THIS DEVICE . ADDITIONALLY, THIS FOLLOW-UP MDR IS BEING SUBMITTED TO CORRECT THE CATALOG NUMBER. IT HAD PREVIOUSLY BEEN REPORTED AS A MEDSTREAM PUMP 20ML; HOWEVER, THE COMPLAINT IS ACTUALLY AGAINST AN UNKNOWN MEDSTREAM REFILL KIT (PRODUCT CODE /LOT NUMBER ARE UNKNOWN). CONCLUSION: BASED ON THE INFORMATION, THE EVENT COULD NOT BE CONFIRMED. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS AND A DEVICE HISTORY RECORD REVIEW CANNOT BE COMPLETED BECAUSE THE LOT FOR THE PRODUCT IS UNKNOWN. SINCE THE EVENT COULD NOT BE CONFIRMED AND THERE IS NO EVIDENCE OF A MANUFACTURING-RELATED MALFUNCTION, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME.

Additional Manufacturer Narrative · 1

THIS FINAL MDR WILL BE THE ONLY REPORT SUBMITTED FOR MFR REPORT #8031062-2017-00011. CONCLUSION: BASED ON THE INFORMATION, THE EVENT COULD NOT BE CONFIRMED. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS AND A DEVICE HISTORY RECORD REVIEW CANNOT BE COMPLETED BECAUSE THE LOT FOR THE PRODUCT IS UNKNOWN. THERE WERE NO SIGNIFICANT SAFETY SIGNALS IDENTIFIED RELATED TO THE REPORTED EVENT BASED ON A REVIEW OF COMPLAINT HISTORY FOR THE DEVICE. SINCE THE EVENT COULD NOT BE CONFIRMED AND THERE IS NO EVIDENCE OF A MANUFACTURING-RELATED MALFUNCTION, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME.

Description of Event or Problem · 1

IN THE LITERATURE ARTICLE ¿INTRATHECAL PUMP REFILLS, POCKET FILLS, AND SYMPTOMS OF DRUG OVERDOSE: A PROSPECTIVE, OBSERVATIONAL STUDY COMPARING THE INJECTED DRUG VOLUME VS. THE DRUG VOLUME EFFECTIVELY MEASURED INSIDE THE PUMP¿ BY PAOLO MAINO, MD; ROBERTO S.G.M. PEREZ, PHD; EVA KOETSIER, MD PUBLISHED NEUROMODULATION: TECHNOLOGY AT THE NEURAL INTERFACE DOI: 10.1111/NER.12597, IT WAS REPORTED THAT PATIENT #1 HAD OVERDOSE SYMPTOMS AFTER MEDSTREAM PUMP REFILL WITH FENTANYL, TWICE, IN (B)(6) 2015. PER THE ARTICLE: ¿ONE COMPLICATION ASSOCIATED WITH THE REFILL PROCEDURE OF INTRATHECAL DRUG DELIVERY SYSTEMS (IDDSS) INCLUDE A ¿POCKET FILL,¿ WHICH IS THE INADVERTENT INJECTION OF THE DRUG INTO THE SUBCUTANEOUS SPACE SURROUNDING THE PUMP. THE PURPOSE OF OUR STUDY WAS TO ASSESS THE MEAN VOLUME DISCREPANCY BETWEEN THE INJECTED AND THE EFFECTIVE DRUG VOLUME INSIDE THE IDDS RESERVOIR AFTER REFILLS THAT DID NOT LEAD TO SIGNS OF OVERDOSE AND TO RECORD AT WHICH VOLUME DISCREPANCY SYMPTOMS OF OVERDOSE WERE NOTICED¿. THE MEDIAN VOLUME DISCREPANCY IN THE 20 ML IDDSS WAS 0.40 ML (IQR: 0.18¿0.61; 10THP: 0.0; 90THP: 0.74) WHEREAS IN THE 40 ML IDDSS THE MEDIAN VOLUME DISCREPANCY WAS 1.3 ML (IQR: 1.06¿1.74; 10THP: 0.81; 90THP: 2.2). SYMPTOMS OF OVERDOSE WERE OBSERVED AFTER 6 OUT OF THE 221 REFILL PROCEDURES. VOLUMES AT WHICH THE SYMPTOMS OF OVERDOSE AROSE WERE BETWEEN 1.15 AND 4.5 ML IN 20 ML IDDSS AND BETWEEN 2.08 AND 4.88 ML IN 40 ML IDDSS¿.. A VOLUME DISCREPANCY OF >1 ML IN 20 ML IDDSS AND OF >2 ML IN 40 ML IDDSS AFTER REFILL SHOULD BE REGARDED AS POTENTIALLY CONSEQUENTIAL WITH REGARD TO RISK OF OVERDOSE POSSIBLY CAUSED BY A POCKET FILL. IN CASE OF POTENTIALLY CONSEQUENTIAL VOLUME DISCREPANCIES, HOSPITALIZATION AND MONITORING IS RECOMMENDED TO PRECLUDE LATE OVERDOSE SYMPTOMS OUTSIDE THE HOSPITAL.¿ THE SYMPTOMS OF OVERDOSE STARTED BETWEEN 5 AND 90 MIN AFTER THE REFILL PROCEDURE. MOST FREQUENT SYMPTOMS OF OVERDOSE WERE SOMNOLENCE, RESPIRATORY DEPRESSION WITH DESATURATION (THE LOWEST VALUE WAS 82%), AND PINPOINT PUPILS. OTHER SYMPTOMS OF OVERDOSE WERE TREMOR, RIGIDITY, AND NAUSEA. IN ALL EXCEPT ONE CASE THE PATIENTS WERE TREATED WITH NALOXONE (BETWEEN 0.2 AND 1.4 MG). THIS LED TO IMMEDIATELY CLINICALLY RECOGNIZABLE ANTAGONIZATION OF THE SYMPTOMS OF OVERDOSE AND WAS ENOUGH TO NORMALIZE THE SPO2 AND RESPIRATORY RATE IN ALL PATIENTS. AS IN 2 PATIENTS THE SYMPTOMS OF OVERDOSE REAPPEARED AFTER ANTAGONIZATION, CONTINUOUS INFUSION OF NAXOLONE WAS STARTED (0.2 MG/ML AT A RATE OF 2¿4 ML/HOUR50.4¿0.8 MG/HOUR) FOR UP TO THREE AND FOUR HOURS, RESPECTIVELY. IN TWO OF THE SIX CASES THE PATIENTS WERE HOSPITALIZED AND MONITORED OVERNIGHT. THE DECISION OF HOSPITALIZATION WAS NOT ONLY TAKEN ON THE BASE OF THE SEVERITY OF THE ADVERSE EVENTS BUT ALSO ON HOME CARE FACILITIES (LIVING ALONE, AGE, ANXIETY OF THE PATIENT). DURING THE HOSPITALIZATION, THESE PATIENTS WERE ALWAYS AROUSABLE, ORIENTED, HEMODYNAMICALLY STABLE, AND MAINTAINED OXYGEN SATURATION LEVELS CONSISTENT WITH THEIR BASELINE. AT THE TIME OF COMPLAINT ENTRY, NO DEVICE SPECIFIC INFORMATION, I.E. CATALOGUE/LOT NUMBER, IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
778215 MEDSTREAM REFILL KIT INFUSION PUMP, IMPLANTABLE, PROGRAMMABLE LKK MEDOS INTERNATIONAL SARL NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention