IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2017-00388
- Event Type
- Malfunction
- Date Received
- November 1, 2017
- Date of Event
- October 3, 2017
- Report Date
- June 21, 2018
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- UDI-DI
- 00613994933713
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
A 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC), SERIAL NUMBER (B)(4): THE DEVICE WAS RETURNED FOR ANALYSIS. ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED. THE FOLLOWING ARE THE TEMPERATURES OF THE DEVICE: AMBIENT TEMPERATURE WAS 78.1 DEGREES F AND MAX TEMPERATURE AFTER 1 MINUTE WAS 81.8 DEGREES F. NO FAULT FOUND WITH THE HANDPIECE. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDICAL DEVICE: PRODUCT ID 1845020 - ATTACHMENT INDIGO ANGLED, SERIAL # (B)(4), LOT # 210760334 MANUFACTURED DATE: APRIL 19, 2016, 510(K) # K081475, UDI (B)(4). PRODUCT ID 1845010 - ATTACHMENT INDIGO STRAIGHT, SERIAL # (B)(4) , LOT # 210812946 MANUFACTURED DATE: FEBRUARY 15, 2016, 510(K) # K081475, UDI (B)(4). NO PRODUCT ANALYSIS RESULTS AVAILABLE. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
CONCOMITANT DEVICES RETURNED FOR ANALYSIS. PRODUCT NUMBER 1845020 ATTACHMENT INDIGO ANGLED, SERIAL NUMBER (B)(4): ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED - TEMPERATURE AFTER 5 MINUTES WAS 97.8 DEGREES F. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. PRODUCT NUMBER 1845010 ATTACHMENT INDIGO STRAIGHT, SERIAL NUMBER (B)(4): ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED - TEMPERATURE AFTER 5 MINUTES WAS 84.1 DEGREES F. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ADDITIONAL INFORMATION RECEIVED INDICATES THE HANDPIECE OVERHEATED IMMEDIATELY (LESS THAN ONE MINUTE) DURING SETUP AND TESTING PRIOR TO USE.
THE CUSTOMER REPORTED THAT THE HANDPIECE WAS HEATING UP, BUR ROTATED SPORADICALLY. THERE WAS NO PATIENT/STAFF IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 774079 | IPC® HANDPIECE - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED INC. | 1845000 | 211085906 | 00613994933713 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |