FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - INDIGO¿ DRILL

MDR report key: 6994487 · Received November 1, 2017

Report

Report Number
1045254-2017-00388
Event Type
Malfunction
Date Received
November 1, 2017
Date of Event
October 3, 2017
Report Date
June 21, 2018
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00613994933713
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC), SERIAL NUMBER (B)(4): THE DEVICE WAS RETURNED FOR ANALYSIS. ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED. THE FOLLOWING ARE THE TEMPERATURES OF THE DEVICE: AMBIENT TEMPERATURE WAS 78.1 DEGREES F AND MAX TEMPERATURE AFTER 1 MINUTE WAS 81.8 DEGREES F. NO FAULT FOUND WITH THE HANDPIECE. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE: PRODUCT ID 1845020 - ATTACHMENT INDIGO ANGLED, SERIAL # (B)(4), LOT # 210760334 MANUFACTURED DATE: APRIL 19, 2016, 510(K) # K081475, UDI (B)(4). PRODUCT ID 1845010 - ATTACHMENT INDIGO STRAIGHT, SERIAL # (B)(4) , LOT # 210812946 MANUFACTURED DATE: FEBRUARY 15, 2016, 510(K) # K081475, UDI (B)(4). NO PRODUCT ANALYSIS RESULTS AVAILABLE. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

CONCOMITANT DEVICES RETURNED FOR ANALYSIS. PRODUCT NUMBER 1845020 ATTACHMENT INDIGO ANGLED, SERIAL NUMBER (B)(4): ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED - TEMPERATURE AFTER 5 MINUTES WAS 97.8 DEGREES F. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. PRODUCT NUMBER 1845010 ATTACHMENT INDIGO STRAIGHT, SERIAL NUMBER (B)(4): ANALYSIS DID NOT CONFIRM THE REPORTED EVENT OF HANDPIECE HEATED UP. PRE-REPAIR TEMPERATURE TESTING WAS PERFORMED - TEMPERATURE AFTER 5 MINUTES WAS 84.1 DEGREES F. THE DEVICE WAS CLEANED AND TESTED TO THE MANUFACTURING SPECIFICATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED INDICATES THE HANDPIECE OVERHEATED IMMEDIATELY (LESS THAN ONE MINUTE) DURING SETUP AND TESTING PRIOR TO USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE HANDPIECE WAS HEATING UP, BUR ROTATED SPORADICALLY. THERE WAS NO PATIENT/STAFF IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
774079 IPC® HANDPIECE - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 211085906 00613994933713

Patients