FDA Adverse Event Injury Summary report: N

EKOSONIC ENDOVASCULAR CATHETER

MDR report key: 6994345 · Received November 1, 2017

Report

Report Number
3001627457-2017-00031
Event Type
Injury
Date Received
November 1, 2017
Date of Event
October 8, 2017
Report Date
October 8, 2017
Manufacturer
EKOS CORPORATION
Product Code
KRA
UDI-DI
00858593006141
PMA / PMN Number
K140151
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED WITH THE SERIAL # CUT OFF THEREFORE THE MANUFACTURE DATE AND EXPIRATION DATE COULD NOT BE IDENTIFIED. THE INVESTIGATION OF THE RETURNED DEVICE ON (B)(6) 2017 CONFIRMED A NICK IN THE OUTER COOLANT WALL. THE LEAK TEST FOR THE MANIFOLD LUER, COOLANT LUER AND DRUG LEUR CONFIRMED NO LEAKS OBSERVED. THE LEAK TEST FOR THE IDDC TUBING CONFIRMED A LEAK DUE TO A NICK IN THE OUTER COOLANT WALL AT 29.5 CM (MEASURING FROM THE DISTAL END OF THE STRAIN RELIEF). IT IS LIKELY THAT THE COMPLAINANT NICKED THE DEVICE WITH A SCALPEL OR SCISSORS. IT SHOULD BE NOTED THAT THE CUT WAS ONLY OBSERVED IN THE COOLANT WALL AND NOT THROUGH THE DRUG LUMEN. BASED ON THE INVESTIGATION, IF THE USER FOLLOWED THE INSTRUCTIONS FOR USE, THE DRUG WOULD HAVE BEEN DELIVERED SUCCESSFULLY THROUGH THE DRUG LUMEN. IF THE USER DELIVERED THE DRUG THROUGH THE COOLANT LUMEN, THIS WOULD EXPLAIN WHY THE PATIENT DID NOT RECEIVE THE TREATMENT AS REPORTED. THE IFU WARNS THE USER THAT IF AN IDDC OR MSD BECOMES KINKED OR OTHERWISE DAMAGED DURING USE, DISCONTINUE USE AND REPLACE. REVIEW OF THE DEVICE HISTORY RECORD INDICATED THAT THE DEVICE WAS MANUFACTURED ACCORDING TO SPECIFICATION AND MET ALL ACCEPTANCE CRITERIA PRIOR TO RELEASE. THE REQUIRED MEDICAL INTERVENTION INCLUDING ADDITIONAL EXPOSURE TO X-RAY AND A NEW CATHETER PLACEMENT WAS DUE TO USER ERROR AND NOT A RESULT OF A DEVICE MALFUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED ON (B)(6) 2017 THAT TWO CATHETERS (12 CM LEFT AND 18 CM RIGHT) WERE PLACED ON (B)(6) 2017 FOR A BILATERAL PE CASE. IT WAS REPORTED THAT ON THE MORNING OF (B)(6) 2017, THE PHYSICIAN DISCOVERED THAT THE RIGHT SIDE CATHETER WAS LEAKING AND IT PREVENTED THE PATIENT FROM RECEIVING TREATMENT IN THE RIGHT PULMONARY ARTERY. THE PATIENT WAS BROUGHT BACK TO THE LAB AND THE PHYSICIAN TOOK A REPEAT ANGIOGRAM OF THE RIGHT PULMONARY ARTERY AND NOTED THAT IT WAS STILL FULL OF THROMBUS. THE PHYSICIAN DOES NOT BELIEVE THAT THE PATIENT RECEIVED ANY TPA IN THE RIGHT PULMONARY ARTERY. IT WAS REPORTED THAT THE PATIENT IS STABLE AND HAS TOLERATED THE TREATMENT TO THIS POINT. THE PLAN IS TO GO ANOTHER 12 HOURS ON THE RIGHT SIDE, OR TREAT SYMPTOMATICALLY DEPENDING UPON THE PATIENT'S RESPONSE. THE CATHETER WAS RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
772672 EKOSONIC ENDOVASCULAR CATHETER CATHETER KRA EKOS CORPORATION 500-55118 00858593006141

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention