FDA Adverse Event Malfunction Summary report: N

EPIDURAL CATHETERIZATION KIT

MDR report key: 6994222 · Received November 1, 2017

Report

Report Number
1036844-2017-00382
Event Type
Malfunction
Date Received
November 1, 2017
Date of Event
October 12, 2017
Report Date
October 12, 2017
Manufacturer
ARROW INTERNATIONAL INC.
Product Code
CAZ
PMA / PMN Number
K103658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION AT THIS TIME. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Additional Manufacturer Narrative · 1

(B)(4). A DEVICE HISTORY RECORD REVIEW WAS PERFORMED ON THE EPIDURAL NEEDLE WITH NO RELEVANT FINDINGS. THE CUSTOMER REPORTED THE EPIDURAL NEEDLE BENT DURING USE. THE CUSTOMER RETURNED ONE EPIDURAL NEEDLE (REFERENCE ATTACHED FILES (B)(4). THE RETURNED NEEDLE WAS VISUALLY EXAMINED WITH AND WITHOUT MAGNIFICATION. VISUAL EXAMINATION OF THE RETURNED NEEDLE REVEALED THE NEEDLE APPEARS USED. THE CANNULA OF THE RETURNED NEEDLE IS BENT. MICROSCOPIC EXAMINATION OF THE NEEDLE BEVEL REVEALED BIOLOGICAL MATERIAL CAN BE SEEN ON THE NEEDLE'S BEVEL. ALSO, THE NEEDLE'S BEVEL TIP IS SLIGHTLY BENT. THE NEEDLE BEVEL APPEARANCE IS SIMILAR TO A NEEDLE BEVEL THAT HAS BEEN PRESSED AGAINST A HARD SURFACE WITH FORCE (REFERENCE FILES (B)(4). A DIMENSIONAL INSPECTION WAS PERFORMED ON THE RETURNED EPIDURAL NEEDLE. THE OUTER DIMENSION (OD) AND INNER DIMENSION (ID) OF THE RETURNED NEEDLE WAS MEASURED. THE OD OF THE RETURNED NEEDLE MEASURED 1.46MM (C05155), WHICH IS WITHIN SPECIFICATION OF 1.46MM-1.48MM PER GRAPHIC NZ-05500-003; REV 7. THE ID MEASURED 0.046IN (1.17MM) (C05157), WHICH IS WITHIN SPECIFICATION OF 1.17MM PER GRAPHIC KZ-05500-007; REV 9. OTHER REMARKS: ADDITIONAL TESTING OF THE NEEDLE TIPS WAS PERFORMED BY THE MANUFACTURING SITE TO DETERMINE TIP STRENGTH. TESTING WAS PERFORMED TO COMPARE NEEDLE TIPS STRENGTHS FROM FIVE AVAILABLE SETS OF DIFFERENT NEEDLES. THE TEST CONCLUDED THAT THE NEEDLE SETS HAD VERY SIMILAR RESULTS. SPECIFICATIONS PER GRAPHIC NZ-05500-003; REV 7 AND KZ-05500-007 REV. 9 WERE REVIEWED AS A PART OF THIS COMPLAINT INVESTIGATION. A REVIEW OF DESIGN CHANGE HISTORY FOR PART NUMBER NZ-05500-001 WAS PERFORMED AS A PART OF THIS INVESTIGATION. NO DESIGN CHANGES HAVE BEEN MADE TO THIS PRODUCT IN THE PAST TWO YEARS THAT WOULD HAVE LED TO THIS COMPLAINT. A REVIEW OF THE QUALITY INSPECTION REPORT FROM THE VENDOR OF THE CANNULA (B)(4) FOUND NO MATERIAL ISSUES FOR LOT # 72P15J0017 (VENDOR LOT # MM150805AJ). THE REPORTED COMPLAINT OF THE NEEDLE BEING BENT DURING USE WAS CONFIRMED BASED ON THE SAMPLE RECEIVED. VISUAL EXAMINATION OF THE RETURNED NEEDLE REVEALED THE CANNULA WAS BENT. ALSO, THE TIP OF THE NEEDLE BEVEL WAS SLIGHTLY BENT, WHICH IS CONSISTENT WITH DAMAGE THAT CAN BE CAUSED WHEN A NEEDLE BEVEL IS PRESSED AGAINST A HARD SURFACE. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED ON THE EPIDURAL NEEDLE WITH NO RELEVANT FINDINGS. NO MATERIAL ISSUES WERE FOUND FROM THE VENDOR FOR THE CANNULA. ALSO, A NEEDLE TIP STRENGTH STUDY WAS PERFORMED BY THE MANUFACTURING SITE WHICH REVEALED VERY SIMILAR RESULTS WITH FIVE DIFFERENT SETS OF NEEDLES INDICATING THAT THE NEEDLE IN QUESTION DOES NOT SHOW ANY MATERIAL DEFICIENCIES OR CHANGES. THEREFORE, BASED UPON THE INFORMATION PROVIDED, THE OBSERVED NEEDLE DAMAGE, AND THE TIME OF DISCOVERY, OPERATIONAL CONTEXT CAUSED OR CONTRIBUTED TO THIS EVENT. HOWEVER, BASED ON THE INCREASED NUMBER OF BENT NEEDLE COMPLAINTS, CAPA REQUEST WAS INITIATED TO FURTHER INVESTIGATE THIS COMPLAINT ISSUE.

Description of Event or Problem · 1

ISSUE REPORTED: THE DOCTOR HAD TWO TUOHY NEEDLES BEND DURING TWO SEPARATE PROCEDURES. THERE WAS NO PATIENT INJURY.

Description of Event or Problem · 1

ISSUE REPORTED: THE DOCTOR HAD TWO TUOHY NEEDLES BEND DURING TWO SEPARATE PROCEDURES. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
773916 EPIDURAL CATHETERIZATION KIT ANESTHESIA CONDUCTION KIT CAZ ARROW INTERNATIONAL INC. 23S16K0528

Patients

Seq Age Sex Outcome Treatment
1