FDA Adverse Event Malfunction Summary report: N

SPECIAL EVD 10-110 W/O Y SITE

MDR report key: 6993074 · Received November 1, 2017

Report

Report Number
2648988-2017-00048
Event Type
Malfunction
Date Received
November 1, 2017
Date of Event
October 3, 2017
Report Date
October 3, 2017
Manufacturer
INTEGRA NEUROSCIENCES PR
Product Code
JXG
PMA / PMN Number
K062599
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION COMPLETED 15DEC2017. TWO SP0090 UNITS WERE RECEIVED; ONE WAS LOT 1170022 AND THE OTHER WAS 1170281. THE ONE FROM LOT 1170281 HAD THE INJECTION PORT DISCONNECTED FROM THE STOPCOCK WHILE 1170022 WAS STILL ATTACHED TO THE STOPCOCK. SINCE 1170022 INJECTION PORT WAS STILL ATTACHED, IT WAS PROCEEDED TO TWIST IT OFF FROM THE STOPCOCK. NO ADHESIVE IS PLACED TO THE INJECTION PORT; THEREFORE, IT CAME OFF AS EXPECTED. DOCUMENTS AND LOGS REVIEWED FOR LOTS 1170022 AND 1170281 DID NOT REVEAL ANYTHING THAT COULD CAUSE THE REPORTED EVENTS. NO DEVIATIONS, EVENTS, AND/OR REWORK WERE RECORDED FOR LOTS 1170022 AND 1170281; THE LOTS COMPLIED WITH ALL REQUIREMENTS AS SPECIFIED IN THE RESPECTIVE MANUFACTURING SHOP ORDER AND RELATED PROCEDURES. LOT NUMBER: 1170022. MANUFACTURING (PACK) DATE: JANUARY 20, 2017. EXPIRATION DATE: DECEMBER 31, 2018 . LOT NUMBER: 1170281. MANUFACTURING (PACK) DATE: FEBRUARY 07, 2017. EXPIRATION DATE: JANUARY 31, 2019. NO DEFECT WAS FOUND ON PRODUCT: THE PRODUCT IS AS PER SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER BELIEVED THAT THERE MAY HAVE BEEN A MANUFACTURING CHANGE ON THE DEVICE. SHE STATED HISTORICALLY, THE PROXIMAL LUER LOCK WAS FUSED TO THE DRAIN AND NOW IT APPEAR TO COME OFF EASILY. THE DRAIN WAS DISCARDED AND REPLACED WITH A NEW ONE. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
774768 SPECIAL EVD 10-110 W/O Y SITE N/A JXG INTEGRA NEUROSCIENCES PR 1170022

Patients

Seq Age Sex Outcome Treatment
1