FDA Adverse Event Injury Summary report: N

AXIOS STENT AND DELIVERY SYSTEM

MDR report key: 6990700 · Received October 31, 2017

Report

Report Number
3005099803-2017-03155
Event Type
Injury
Date Received
October 31, 2017
Date of Event
September 21, 2017
Report Date
October 5, 2017
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
PCU
PMA / PMN Number
K150692
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINANT INDICATED THAT THE DEVICE HAS BEEN DISPOSED OF AND WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED, A SUPPLEMENTAL MDR WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AXIOS STENT WAS TO BE IMPLANTED TRANSDUODENAL TO THE COMMON BILE DUCT DURING AN ENDOSCOPIC ULTRASOUND (EUS)-GUIDED BILIARY DRAINAGE PROCEDURE ON (B)(6) 2017. REPORTEDLY, THE PATIENT HAD PANCREATIC CANCER (LIKELY PANCREATIC ADENOCARCINOMA, LOCALLY ADVANCED) AND AN OBSTRUCTED DUODENUM. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, EUS AND FLUOROSCOPY WERE USED AND THE SCOPE WAS IN A VERY TORTUOUS AND DIFFICULT POSITION TO ACCESS THE BILE DUCT. THE PHYSICIAN USED A GUIDEWIRE DURING THE CASE AND WAS UNSURE OF ITS ACCURATE POSITION. AFTER DEPLOYING THE AXIOS STENT, THE PHYSICIAN REALIZED THAT IT HAD BEEN DEPLOYED IN THE INCORRECT POSITION, WITH ONE FLANGE DEPLOYED IN THE BILE DUCT AND ONE FLANGE DEPLOYED IN THE SPACE BETWEEN THE DUODENAL WALL AND THE WALL OF THE BILE DUCT. THE PATIENT WAS TAKEN TO SURGERY, WHERE THEY UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY AND EXPLORATORY LAPAROSCOPY. THE PHYSICIAN REPORTED THAT THE AIM OF THE LAPAROSCOPIC CHOLECYSTECTOMY WAS TO ATTEMPT TO PASS A TRANSCYSTIC STENT TO SEAL THE AXIOS PUNCTURE SITE IN THE COMMON BILE DUCT WALL. THE STENT WAS REMOVED FROM THE PATIENT DURING THE LAPAROSCOPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
770119 AXIOS STENT AND DELIVERY SYSTEM PANCREATIC STENT, COVERED, METALLIC, REMOVABLE PCU BOSTON SCIENTIFIC - GALWAY M00553520 0020649762

Patients

Seq Age Sex Outcome Treatment
1 92 YR Required Intervention